2620A, 2625A

Users Manual

Table 4-8. Command and Query Reference (cont)

LOGGED?

LOG_BIN?

LOG_CLR

LOG_COUNT?

The Totalizer count is returned as a scientific notation value in the range 0 through 65535 (00.000E+3 through 65.535E+3). If the Totalizer has overflowed, a value of 1E+9 is returned.

An Execution Error is generated if this query is used with the IEEE-488 interface.

The following example shows the type of data received in response to LOG?. The first line shows time and date. The second line shows measurement values. The third line shows alarm outputs state, DIO state, and totalizer value. Note that the actual response combines all this information on one line; three lines are shown here for presentation clarity.

16,15,30,10,3,90,

-034.53E-3,+09.433E+0,+09.433E+0,+09.434E+0,+09.434E+0, 15,255,+00.000E+3

Retrieve specified scan data from memory (RS-232 only)

LOGGED? <index>

<index> = (1 .. 2047)

Scan data is not removed from memory with this query. Use of <index> values outside the range causes an Execution Error. Use of <index> values within the range and have no data stored will also cause an Execution Error. An Execution Error is generated if the instrument is a Hydra Data Acquisition Unit.

Scan data is returned in the same format as for the LOG? query.

Binary Upload of Logged Data (RS-232 only)

For quick upload of logged data from a 2625A. See Appendix E. An Execution Error is generated if the instrument is a Hydra Data Acquisition Unit, or <index> is out of range or invalid.

LOG_BIN? <index>

Clear Logged Scans (RS-232 only)

Clear all stored scan data. An Execution Error is generated if the instrument is a Hydra Data Acquisition Unit.

Logged Scan Count Query (RS-232 only)

Return the number of stored scans. Returns an integer value representing the number of scans presently stored in memory. The maximum value that can be returned is 2047; 0 indicates that there are no stored scans. An Execution Error is generated if the instrument is a Hydra Data Acquisition Unit.

4-36

Page 106
Image 106