Fluke 2620A, 2625A Maintenance, Performance Tests, 1. Power-Up Error Codes Description

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 133
Image 133

Maintenance 6

Performance Tests

Error

1

2

3

4

5

6

7

8

9 A b

C

Table 6-1. Power-Up Error Codes

Description

ROM checksum error

External RAM test failed

Internal RAM test failed

Display power-up test failure

Display not responding

Instrument configuration corrupted

EEPROM instrument configuration corrupted EEPROM calibration data corrupted

A/D not responding

A/D ROM test failure

A/D RAM test failure

A/D self test failure

Each of the measurements listed in the following steps assume the instrument is being tested after a 1/2 hour warm up, in an environment with an ambient temperature of 18 to 28 degrees C, and a relative humidity of less than 70%.

Note

All measurements listed in the performance test tables are made in the slow reading rate unless otherwise noted.

Warning

The 2620A/2625A instrument contains high voltages that can be dangerous or fatal. Only qualified personnel should attempt to service the instrument. Turn off the Hydra Series II and remove all power sources before performing the following procedures.