Fluke 2625A, 2620A user manual Front Panel Lock out Conditions, Front Panel Review Only Function

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 68
Image 68

2620A, 2625A

Users Manual

Table 3-17. Autoprint/Memory Storage Selection

Press these buttons:P

To select from these choices:

MODE

KU)

Destination (“dESt”) (Note 1)

G

D

Print

StorE

both

E

Mode (“MOdE”) (Note 2)

G

D

ALL

ALAr trAnS

E

Note 1.

“Print”

Sends data to be printed through the RS-232 interface.

 

“StorE”

Sends data to be stored in memory. (Hydra Data Logger only.)

 

 

 

 

 

 

“both”

Sends data to be stored and printed (Hydra Data Logger only.)

Note 2.

“ALL”

Measurements for all defined channels are printed/stored when a scan occurs.

 

 

 

 

 

“ALAr”

Measurements for all defined channels are printed/stored when a scan occurs

 

 

and at least one channel is in alarm.

 

“trAnS”

Measurements for all defined channels are printed/stored when a scan occurs

 

 

 

and at least one channel has transitioned into or out of alarm since the last

 

 

scan.

 

 

 

 

 

 

 

Table 3-18. Clearing Memory Storage

To Initiate

Hold K, and press U

To Confirm

“no” to exit without clearing

G D

To Activate

E

“CLEAr” appears in right display

“YES”

Cto exit without clearing

Front Panel Lock out Conditions

Various methods are available to prevent accidental use of the front panel buttons. These actions can be initiated from either the front panel or the computer interface.

Front Panel Review Only Function

Access the Review Array (press N, REVIEW annunciator comes on)

Press Fand Bsimultaneously to activate the Review Only function (REM annunciator comes dim).

J, H, G, and Dcan now be used to view the various elements of the Review Array. All other front panel buttons are locked out, yielding a long beep when pressed. Press both Fand Bagain to deactivate the Review Only function and return the instrument to normal front panel button operation (regular Review Array display.)

Front Panel Monitor Only Function

First place the instrument in the Monitor Function (press M, MON annunciator comes on). Then press Fand Bsimultaneously to activate the Monitor Only Function (REM annunciator comes on).

3-26

Page 68
Image 68
Fluke 2625A, 2620A Front Panel Lock out Conditions, Front Panel Review Only Function, Front Panel Monitor Only Function