Specifications A

Ohms Inputs

Range

Resolution

 

 

 

 

Slow

Fast

 

 

 

300 Ω

10 mΩ

0.1 Ω

3 kΩ

0.1 Ω

1 Ω

30 kΩ

1 Ω

10 Ω

300 kΩ

10 Ω

100 Ω

3 MΩ

100 Ω

1 kΩ

10 MΩ

1 kΩ

10 kΩ

 

 

 

Typical Full

Scale Voltage

0.22V 0.25V

0.29V

0.68V

2.25V

2.72V

Maximum

Current

Through

Unknown

1mA 110 μA 13 μA 3.2 .μA 3.2 μA 3.2 μA

Maximum Open

Circuit Voltage

3.2V

1.5V

1.5V

3.2V

3.2V

3.2V

Range

 

4-Wire Accuracy ±(%±Ω)

 

 

 

 

 

 

 

18°C to 28°C

 

0°C to 60°C

 

 

 

 

 

90 Days, Slow

1 Year, Slow

1 Year, Fast

1 Year, Slow

1 Year, Fast

 

 

 

 

 

300Ω

3 kΩ

30 kΩ

300 kΩ

3 MΩ

10 MΩ

0.013% + 20 mΩ

0.015% + 0.2 Ω

0.013% + 2 Ω

0.020% + 20 Ω

0.059% + 200 Ω

0.168% + 2 kΩ

0.014% + 20 mΩ

0.016% + 0.2 Ω

0.014% + 2 Ω

0.021% + 20 Ω

0.063% + 200 Ω

0.169% + 2 kΩ

0.014% + 0.2 Ω

0.016% + 2 Ω

0.014% + 20 Ω

0.021% + 200 Ω

0.063% + 2 kΩ

0.709% + 20 kΩ

0.031% + 20 mΩ

0.039% + 0.2 Ω

0.039% + 2 Ω

0.050% + 20 Ω

0.231% + 200 Ω

0.573% + 2 kΩ

0.031% + 0.2 Ω

0.039% + 2 Ω

0.039% + 20 Ω

0.050% + 200 Ω

0.231% + 2 kΩ

0.923% + 20 kΩ

2-Wire Accuracy

Add 4.0Ω to accuracy specifications for channels 1 to 20, and add 20 mΩ for channel 0. Lead wire resistances are not included.