Fluke 2620A General, Appendices, Altitude, Vibration, Shock, Channel Capacity, Measurement Speed

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 167
Image 167

Appendices

Specifications A

Altitude

Operating: 2,000 m maximum

Non-operating: 12,200 m maximum

Vibration

0.7g at 15 Hz

1.3g at 25 Hz 3 g at 55 Hz

Shock

30g half sine per Mil-T-28800 Bench handling per Mil-T-28800

General

Channel Capacity

21 Analog Inputs

4 Alarm Outputs

8 Digital I/O (inputs/outputs)

Measurement Speed

Slow rate: 4 readings/second nominal

Fast rate: 17 readings/second nominal

1.5 readings/second nominal for ACV and high-Ωinputs

For additional information, refer to Typical Scanning Rated and Maximum Autoranging Time.

Memory Life

10 years minimum over Operating Temperature range

Stores: real-time clock, set-up configuration, and measurement data

Common Mode Voltage

300V dc or ac rms maximum from any analog input (channel) to earth provided that channel to channel maximum voltage ratings are observed.

Voltage Ratings

Channels 0,1, and 11 are rated at 300V dc or ac rms maximum from a channel terminal to earth and from a channel terminal to any other channel terminal.

Channels 2 to 10 and 12 to 20 are rated at 150V dc or ac rms maximum from a channel terminal to any other channel terminal within channels 2 to 10 and 12 to 20.

IEC Overvoltage Category II.

Size

9.3 cm high, 21.6 cm wide, 31.2 cm deep

Weight

Net, 2.95 kg

Shipping, 4.0 kg

A-15