Fluke user manual AC Voltage Inputs, 2620A/2625A, Wire Accuracy, Maximum Current Through Sensor

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 158
Image 158

2620A/2625A

Users Manual

IEC 751, Amendment 2, 100 e Platinum (ITS-90)

RTD

 

 

 

4 Wire Accuracy* (±°C)

 

 

 

 

 

 

 

 

Temperature

Resolution

 

 

18°C to 28°C

 

 

 

 

 

 

 

 

 

(°C)

Slow

Fast

90 Day

 

1 Year

 

1 Year

 

 

 

Slow

 

Slow

 

Fast

 

 

 

 

 

 

 

 

0°C to 60°C

1 Year

1 Year

Slow

Fast

 

 

-200.00

0.02

0.1

0.13

0.00

 

 

 

0.02

0.1

0.09

100.00

 

 

 

0.02

0.1

0.12

300.00

 

 

 

0.02

0.1

0.23

600.00

 

 

 

0.02

0.1

0.55

 

 

 

 

* Sensor inaccuracies are not included

0.13

0.09

0.12

0.24

0.56

0.54

0.55

0.59

0.74

1.12

0.130.55

0.130.59

0.180.66

0.340.85

0.731.29

2-Wire Accuracy

For 2-wire sensors with R0= 100Ω: degrade accuracy by 5.0ºC per lead-ohm, plus degrade accuracy an additional 11ºC for channels 1 to 20 and 0.05ºC for channel 0.

Maximum Current Through Sensor

1 mA

Typical Full Scale Voltage

0.22 V

Maximum Open Circuit Voltage

3.2 V

Maximum Sensor Temperature

600ºC nominal

999.99ºF max displayed

Cross-talk Rejection

Refer to Appendix D.

AC Voltage Inputs

Range

 

Resolution

 

 

 

 

 

Slow

 

Fast

 

 

 

 

300 mV

10 μV

 

100 μV

3 V

100 μV

 

1 mV

 

30 V

1 mV

 

10 mV

 

300 V

10 mV

 

100 mV

 

 

 

 

 

Minimum Input For

Rated Accuracy

20mV 200 mV

2 V

20 V

A-6

Page 158
Image 158
Fluke AC Voltage Inputs, 2620A/2625A, Wire Accuracy, Maximum Current Through Sensor, Typical Full Scale Voltage