1.TECHNICAL DATA

1.1General Safety Information

Maintenance and repair may be performed only by an office that is authorized by the manufacturer.

In the Federal Republic of Germany, electrical installation of medically used rooms must conform to VDE Regulation 0107. In all other countries, the particular applicable national regulations take precedence and must be observed. These can be found in the system project plan.

During installation, it must be assured that all ground wire connections provided by the manufacturer must be connected prior to starting up the unit.

The ground wire between the individual components and the power supply must be connected as shown in the Wiring Diagram.

The regulations provided by the trade unions for occupational safety and prevention of accidents must be observed.

No work may be performed on parts that are under power (above 42 V).

This condition applies for both measurement and adjustment work steps. However, special care is required when performing these worksteps.

If the operating instructions call for voltage to be switched on for unit movements, the system must be shut down immediately following such tests.

The radiation regulations must be observed while making settings and checks that must be performed under X-radiation; radiation protective clothing must be worn.

1.2 Component Weights

Unit column

approx. 320 kg

Table

approx. 230 kg

Tube unit support arm

approx.

50 kg

Footswitch

approx.

5 kg

Total weight, incl. tube unit, collimator,

 

 

adjustable Bucky and I.I.:

approx. 730 kg

1.3 Component Dimensions

Unit column with table and tube unit support arm on the pallet.

2050 mm x 1015 x 1800 mm

1.4 Packaging Dimensions and Weights

Unit column with table, incl. accessories

2050 mm x 1015 mm x 1890 mm

 

approx. 900 kg

 

Transport carriage

1800 mm x 200 mm x

200 mm

 

approx. 85 kg

 

0116 7201

-4 of 90 -

06/00

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Image 4
Contents Installation Instructions Uromat Contents Technical Maintenance AdjustmentsNot configured General Safety Information Component WeightsComponent Dimensions Packaging Dimensions and WeightsComponent Designations Dimensional Drawing, Right-handed Version FFADimensional Drawing, Left-handed Version FFA345 Drehachse Dimensional Drawing for Floor Mounting, Left-handed Version 0116 06/00 Power Line Connection Data Required Special ToolsRequired Test Equipment Conditions for Transport and StorageCircuit Diagram, Contents of Circuit Diagram06/00 0116 Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Collimator connection System Location Component Function Component Designations for the Circuit DiagramUromat LED1 +A6 Terminal strip for Bucky Cassette box motor potentiometer Function Description of the Controller Boards CPU BoardBreaker Board Footswitch Board Control Unit Display BoardPreparations for Mounting Unpacking the UnitRemoving the Unit from the Pallet Installing the Unit Column Aligning the Unit Column Temporarily Connecting Line Power Installing the Tube Unit Support ArmPreinstalling the Tube Unit Collimator Combination Installing the Image Intensifier Laying the Collimator CablesLaying the High Voltage Cables To lay cables in the unit, 7 meters are requiredPage Making the Power Connection through the Generator Laying the I.I. CableInstalling the Measuring Chamber Installing the Grid Replacing the Grid Installing the Tilt Angle Display- Error DisplayInstalling the Cover Panels Installing the Monitor Support Arm OptionInstalling the Mount for the Flush Bowl Sealing the Table Frame Cover PanelAdjustment Tube Unit Support ArmCentral Beam Center Central Beam Center of BuckyLengthwise to the tabletop, for this Movement Path Potentiometer, Tabletop longitudinal Bucky Cassette Tray Movement Path Potentiometer Movement Path Potentiometer, Tabletop transverseI.I. Carriage System Potentiometer Potentiometer 1 Table Tilt Movement Path Potentiometer 2 Table Tilt Movement PathTabletop Longitudinal Safety Switch Tabletop Transverse Safety Switch Not configuredBucky Movement Path Safety Switch 13 I.I. Carriage Movement Path Safety SwitchTable Up Movement Path Switch Strike Plate Table Down Movement Path Switch Strike Plate17 +88 Tilt Movement Safety Switch 16 -20 Tilt Movement Safety SwitchSpindle Nut Safety Limit Switch Tabletop longitudinal Belt Tension Accessory Sensor SwitchesRaise-Lower Tabletop Belt Tension Cassette Carriage Drive Belt Tension Operation in the Service Mode EquipmentSetting Parameters ProcedureAdjusting the Axes StartupCollimator Basic Setting Cassette Format Setting the Collimator to the Cassette and I.I. FormatFormat Saving the Parameters in the PC Electronic Adjustment Parameters Setup ParametersTabletop long. Store yes/no 1/0 Technical Maintenance Mechanical and Electrical ChecksPreparations Page Function Check Page 176 Spare Parts, Overview41/43 40/41 /10 Spare Parts, Unit Column Right-handed, Left-handed Versions122 125 11 119 120 121 151 Spare Parts, Table Frame Tabletop, Right-handed Version128 135 185 136 Spare Parts, Table Frame Tabletop, Left-handed VersionSpare Parts, Tube Unit Support Right-handed Version 193 191 213 209 202 203 191 192 197 194 193 198 201 207 / 210209 207 / 210 Spare Parts, Tube Unit Support Left-handed VersionSpare Parts, Accessories 249 155 156Spare Parts List Pos DesignationPos Designation Pos Designation Pos Designation Pos Designation Troubleshooting Error Cause Corrective Measure NumberPage Board Replacement Version UpdateMaintenance Verification Location of Identification Labels Labeling Max kg/ 66 lb Overview of Labels