Tabletop transverse movement:
-Even movement?
-Soft startup and soft outward movement?
-Automatic stop in the centered position?
-Automatic stop in the limit position?
-Movement initiated from footswitch?
-Movement initiated from the manual control unit?
Movable Tube Unit Support Arm:
-Easy separation in the exposure position?
-Easy separation in the park position?
-Even movement?
-Positive engagement into the limit positions?
Table extension:
-Does the table extension engage in place in the opening at the foot end?
-Is the table extension blocked from sliding out?
-Does the controller when the table extension is controlled?
Footrest:
- Does the footrest securely and easily engage the footrest?
Micturation seat:
-Does the micturation seat engage in place in the opening?
-Is the micturation seat secured against falling out?
-Does the controller sense the micturation seat?
Accessories:
- Are all configured accessories functional?
General:
-Is the centering of the tube unit to the adjustable Bucky okay?
-Is the centering of the tube unit to the I.I. okay?
-Is radiation blocked when the tube unit support arm is in the park position?
06/00 | - 71 of 90 - | 0116 7201 |
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf