Hansol Multitech Uromat 3000 Setting the Collimator to the Cassette and I.I. Format

Page 63

3.26 Setting the Collimator to the Cassette and I.I. Format

Cassette Format:

To adjust the collimator to the 4 possible film formats, the corresponding cassette must be inserted.

Use the button (Fig. 78/Pos. 33) to switch on the collimator light field and center the cassette template included in the shipment so it is centered on the tabletop.

With the light field switched on, move the collimator plates using the buttons (Fig. 78/Pos. 31, 32, 35, 37) horizontal open, horizontal closed, vertical open, vertical closed to the position of the cassette template and save the position with the Store (Pos. 54) and Light (Pos. 33) buttons.

Then as a control, make a radiographic exposure and, if needed, correct the collimator plate adjustment.

Caution:

During exposure, radiation protective measures must be observed.

Perform this routine with all 4 cassettes.

A format that has been saved previously can be overwritten by repeating the save routine.

I.I. Format:

The unit can store 3 Zoom steps.

LED`s Z-N (Pos. 41) = I.I. size, LED`s Z-1 (Pos. 42) = Zoom step 1, LED`s Z-2 (Pos. 43) = Zoom step 2, LED`s Z-1 + Z-2 (Pos. 41 + 42) = Zoom step 3.

The Zoom step is changed by pressing the Zoom button (Pos. 44).

Use the button (Fig. 78/Pos. 33) to switch on the collimator light field and center the cassette template included in the shipment on the tabletop.

Move the iris diaphragm to the desired position with the keys Iris open (Fig. 78/Pos. 38) or Iris closed (Pos. 36).

To save the currently selected zoom step, press and hold the Store button (Pos. 54) on the manual controller and then, both iris buttons (Pos. 36 and 38) should be pressed either together or in sequence.

Then the adjustment should be checked, and if needed, corrected under fluoroscopy.

Caution:

During radiographic fluoroscopy, radiation safety regulations must be observed.

Since the position of the horizontal and vertical collimator plates also need to be saved for improved radiation safety, they must be moved as exactly as possible to the limit of the iris diaphragm without them being visible in the radiation field.

Save the position of the collimator plates with the Store (Fig. 78/Pos. 54) and Light (Pos. 33) buttons.

Perform this routine for all zoom steps.

The values can be resaved as often as wished, the old settings will be overwritten when this is done.

After completing adjustment work, all parameters must be saved in the EEprom by pressing the Reset button (Pos. 56).

The switch the unit into the normal mode with the switch (Fig. 77/Pos. 1), the green LED (Pos. 2) on the controller blinks.

06/00

- 63 of 90 -

0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Image 63
Contents Installation Instructions Uromat Contents Adjustments Technical MaintenanceNot configured Packaging Dimensions and Weights General Safety InformationComponent Weights Component DimensionsComponent Designations FFA Dimensional Drawing, Right-handed VersionFFA Dimensional Drawing, Left-handed Version345 Drehachse Dimensional Drawing for Floor Mounting, Left-handed Version 0116 06/00 Conditions for Transport and Storage Power Line Connection DataRequired Special Tools Required Test EquipmentContents of Circuit Diagram Circuit Diagram,06/00 0116 Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Collimator connection Component Designations for the Circuit Diagram System Location Component FunctionUromat LED1 +A6 Terminal strip for Bucky Cassette box motor potentiometer Function Description of the Controller CPU Board BoardsBreaker Board Display Board Footswitch Board Control UnitUnpacking the Unit Preparations for MountingRemoving the Unit from the Pallet Installing the Unit Column Aligning the Unit Column Installing the Tube Unit Support Arm Temporarily Connecting Line PowerPreinstalling the Tube Unit Collimator Combination Laying the Collimator Cables Installing the Image IntensifierTo lay cables in the unit, 7 meters are required Laying the High Voltage CablesPage Laying the I.I. Cable Making the Power Connection through the GeneratorInstalling the Measuring Chamber Installing the Tilt Angle Display- Error Display Installing the Grid Replacing the GridInstalling the Monitor Support Arm Option Installing the Cover PanelsSealing the Table Frame Cover Panel Installing the Mount for the Flush BowlTube Unit Support Arm AdjustmentCentral Beam Center of Bucky Central Beam CenterLengthwise to the tabletop, for this Movement Path Potentiometer, Tabletop longitudinal Movement Path Potentiometer, Tabletop transverse Bucky Cassette Tray Movement Path PotentiometerI.I. Carriage System Potentiometer Potentiometer 2 Table Tilt Movement Path Potentiometer 1 Table Tilt Movement PathTabletop Transverse Safety Switch Not configured Tabletop Longitudinal Safety SwitchTable Down Movement Path Switch Strike Plate Bucky Movement Path Safety Switch13 I.I. Carriage Movement Path Safety Switch Table Up Movement Path Switch Strike Plate16 -20 Tilt Movement Safety Switch 17 +88 Tilt Movement Safety SwitchSpindle Nut Safety Limit Switch Accessory Sensor Switches Tabletop longitudinal Belt TensionRaise-Lower Tabletop Belt Tension Cassette Carriage Drive Belt Tension Equipment Operation in the Service ModeProcedure Setting ParametersStartup Adjusting the AxesCollimator Basic Setting Setting the Collimator to the Cassette and I.I. Format Cassette FormatFormat Saving the Parameters in the PC Setup Parameters Electronic Adjustment ParametersTabletop long. Store yes/no 1/0 Mechanical and Electrical Checks Technical MaintenancePreparations Page Function Check Page Spare Parts, Overview 176Spare Parts, Unit Column Right-handed, Left-handed Versions 41/43 40/41 /10Spare Parts, Table Frame Tabletop, Right-handed Version 122 125 11 119 120 121 151Spare Parts, Table Frame Tabletop, Left-handed Version 128 135 185 136213 209 202 203 191 192 197 194 193 198 201 207 / 210 Spare Parts, Tube Unit Support Right-handed Version 193 191Spare Parts, Tube Unit Support Left-handed Version 209 207 / 210249 155 156 Spare Parts, AccessoriesPos Designation Spare Parts ListPos Designation Pos Designation Pos Designation Pos Designation Error Cause Corrective Measure Number TroubleshootingPage Version Update Board ReplacementMaintenance Verification Location of Identification Labels Labeling Overview of Labels Max kg/ 66 lb