3.23 Operation in the Service Mode
a. Connection of a Laptop
To establish a connection with the URO controller, a PC/laptop is required.
Equipment:
⇒PC or laptop with color display (black/white is also sufficient, but makes operation more difficult)
⇒486 processor, 16 MB RAM
⇒Windows 3.x Win 95 operating system (Windows NT can cause difficulties with the Com interface)
⇒Hard drive
⇒Serial connection cable
⇒Current version of the
b. Procedure
∙Switch off the URO
∙Connect the PC and the URO CPU board using the connection cable
∙Switch on the URO
∙After approx. 1 s, the service LED will begin to blink
∙Switch on the PC
∙Start the PC software
∙Set the PC software to monitor. The terminal screen will appear
∙Set the service switch on the CPU board to the up "Service" position
∙After approx. 1 s, the service LED will switch to continuously on
∙The start message of the service menu will appear on the screen.
U | U | RRRR | OOO | 333 | 000 | 000 | 000 |
U U | R R | O O |
| 3 0 00 0 00 0 00 | |||
U | U | RRRR | O O | 33 | 0 0 0 0 0 0 0 0 0 | ||
U U | R R | O O |
| 3 00 0 00 0 00 0 | |||
| UUU | R R | OOO | 333 | 000 | 000 | 000 |
|
|
| Service mode |
|
|
| |
|
| (1) | Input test |
|
|
| |
|
| (2) | Analog test |
|
|
| |
|
| (3) | Handcontrol and footswitch test | ||||
|
| (4) | Axis calibration |
|
| ||
|
| (5) | Versioncheck |
|
|
Select testmode
06/00 | - 59 of 90 - | 0116 7201 |
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf