F 40 | Format switch cassette incorrect. | Check 4 format switches. | |
F 41 | M1 motor breaker stuck/defective. | Check K1 breaker. | |
F 42 | M2 motor breaker stuck /defective. | Check K2 breaker. | |
F 43 | M3 motor breaker stuck /defective. | Check K3 breaker. | |
F 44 | M4 motor breaker stuck /defective. | Check K4 breaker. | |
F 45 | M5 motor breaker stuck /defective. | Check K5 breaker. | |
F 46 | M6 motor relay stuck /defective. | Check K19 relay (CPU board). | |
F 47 | FU0 N>0 active. | FU0 relay defective/ input | |
|
|
| defective / check cabling. |
F 48 | FU1 N>0 active. | FU1 relay defective / input | |
|
|
| defective / check cabling. |
F 50 | Collimator malfunction. | Check cabling, potentiometers, | |
|
|
| setup parameters. |
|
|
| |
F 61 | Limit switch, temperature, breaker | Free run (press button on breaker | |
| malfunction, M1 tilt /vertical drive or M3. | board). Allow motor to cool down. | |
F 62 | Limit switch, temperature, breaker | Free run (press button on breaker | |
| malfunction, M2 system drive. | board). Allow motor to cool down. | |
F 64 | Limit switch, temperature, breaker | Free run (press button on breaker | |
| malfunction, M4 table longitudinal drive. | board). Allow motor to cool down. | |
F 79 | Invalid parameter version | Initialize correct parameter | |
|
|
| version. |
F 80 | EEPROM memory error. | Download new parameter file. | |
|
|
| |
|
|
| |
F 99 | Unit is in the service mode. | Set the service switch to the RUN | |
|
|
| mode. |
F = Fatal Error | = System error = cannot be corrected |
06/00 | - 85 of 90 - | 0116 7201 |