Hansol Multitech Uromat 3000 Troubleshooting, Error Cause Corrective Measure Number

Page 84

5.Troubleshooting

Error

Cause

Corrective Measure

Number

 

 

F

1

Frequency converter 0 for tilt/vertical,

Check connection cables, settings

 

 

system failed.

(address, Baud rate).

F

2

Frequency converter 1 for tilt/vertical, table

Check connection cables, settings

 

 

longitudinal movement, cassette failed.

(address, Baud rate).

F

3

Frequency converters 0+1 failed.

Check connection cables, settings

 

 

 

(address, Baud rate).

 

 

 

 

 

 

 

 

 

 

 

 

F 10

Footswitch communication problem.

Check fuse, cabling.

F 11

Manual control unit communication

Check fuse, cabling.

 

 

problem.

 

 

 

 

F 13

Incorrect footswitch software version.

Replace footswitch.

F 14

Incorrect manual control unit software

Replace manual control unit.

 

 

version.

 

 

 

 

 

 

 

 

 

 

F 20

Analog voltage failed.

Check fuse.

F 21

M1 tilt/vertical potentiometer error.

Check potentiometer cables and

 

 

 

potentiometer.

F 22

System potentiometer error.

Check potentiometer cables and

 

 

 

potentiometer.

F 23

M3 tilt/vertical potentiometer error.

Check potentiometer cables and

 

 

 

potentiometer.

F 24

Table longitudinal potentiometer error.

Check potentiometer cables and

 

 

 

potentiometer.

F 25

Cassette potentiometer error.

Check potentiometer cables and

 

 

 

potentiometer.

F26

Table transverse potentiometer error /

Check motor potentiometer cables

 

 

motor defective.

and potentiometer.

 

 

 

F 28

Collimator vertical potentiometer error.

Check potentiometer cables and

 

 

 

potentiometer.

 

 

 

F 30

Button fuse malfunction.

Check buttons, cables.

F 31

Footswitch button sticking.

Check buttons.

F 32

Manual control unit button sticking.

Check buttons.

F 33

Exposure switch active during system start.

Check switches.

F 34

Fluoroscopy switch active during system

Check switches.

 

 

start.

 

F 35

Exposure centering switch defective.

Check switches, cabling.

 

 

 

 

 

 

 

 

 

0116 7201

-84 of 90 -

06/00

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Image 84
Contents Installation Instructions Uromat Contents Adjustments Technical MaintenanceNot configured General Safety Information Component WeightsComponent Dimensions Packaging Dimensions and WeightsComponent Designations Dimensional Drawing, Right-handed Version FFADimensional Drawing, Left-handed Version FFA345 Drehachse Dimensional Drawing for Floor Mounting, Left-handed Version 0116 06/00 Power Line Connection Data Required Special ToolsRequired Test Equipment Conditions for Transport and StorageCircuit Diagram, Contents of Circuit Diagram06/00 0116 Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Circuit Diagram, Collimator connection Component Designations for the Circuit Diagram System Location Component FunctionUromat LED1 +A6 Terminal strip for Bucky Cassette box motor potentiometer Function Description of the Controller Boards CPU BoardBreaker Board Footswitch Board Control Unit Display BoardPreparations for Mounting Unpacking the UnitRemoving the Unit from the Pallet Installing the Unit Column Aligning the Unit Column Temporarily Connecting Line Power Installing the Tube Unit Support ArmPreinstalling the Tube Unit Collimator Combination Installing the Image Intensifier Laying the Collimator CablesLaying the High Voltage Cables To lay cables in the unit, 7 meters are requiredPage Making the Power Connection through the Generator Laying the I.I. CableInstalling the Measuring Chamber Installing the Grid Replacing the Grid Installing the Tilt Angle Display- Error DisplayInstalling the Cover Panels Installing the Monitor Support Arm OptionInstalling the Mount for the Flush Bowl Sealing the Table Frame Cover PanelAdjustment Tube Unit Support ArmCentral Beam Center of Bucky Central Beam CenterLengthwise to the tabletop, for this Movement Path Potentiometer, Tabletop longitudinal Movement Path Potentiometer, Tabletop transverse Bucky Cassette Tray Movement Path PotentiometerI.I. Carriage System Potentiometer Potentiometer 1 Table Tilt Movement Path Potentiometer 2 Table Tilt Movement PathTabletop Longitudinal Safety Switch Tabletop Transverse Safety Switch Not configuredBucky Movement Path Safety Switch 13 I.I. Carriage Movement Path Safety SwitchTable Up Movement Path Switch Strike Plate Table Down Movement Path Switch Strike Plate16 -20 Tilt Movement Safety Switch 17 +88 Tilt Movement Safety SwitchSpindle Nut Safety Limit Switch Accessory Sensor Switches Tabletop longitudinal Belt TensionRaise-Lower Tabletop Belt Tension Cassette Carriage Drive Belt Tension Operation in the Service Mode EquipmentSetting Parameters ProcedureAdjusting the Axes StartupCollimator Basic Setting Setting the Collimator to the Cassette and I.I. Format Cassette FormatFormat Saving the Parameters in the PC Electronic Adjustment Parameters Setup ParametersTabletop long. Store yes/no 1/0 Technical Maintenance Mechanical and Electrical ChecksPreparations Page Function Check Page 176 Spare Parts, Overview41/43 40/41 /10 Spare Parts, Unit Column Right-handed, Left-handed Versions122 125 11 119 120 121 151 Spare Parts, Table Frame Tabletop, Right-handed Version128 135 185 136 Spare Parts, Table Frame Tabletop, Left-handed VersionSpare Parts, Tube Unit Support Right-handed Version 193 191 213 209 202 203 191 192 197 194 193 198 201 207 / 210209 207 / 210 Spare Parts, Tube Unit Support Left-handed VersionSpare Parts, Accessories 249 155 156Spare Parts List Pos DesignationPos Designation Pos Designation Pos Designation Pos Designation Troubleshooting Error Cause Corrective Measure NumberPage Board Replacement Version UpdateMaintenance Verification Location of Identification Labels Labeling Max kg/ 66 lb Overview of Labels