-Check automatic stop in
-Clean the guide rails.
-Check for play in the bearings. If necessary, readjust.
Tube unit support arm, exposure position, park position:
-Press the buttons on the control handle and move through the entire movement range.
-Check for even movement during the unit movements.
-Check for engagement into the end positions.
-Clean the guide rails.
-Check for play in the bearings. If necessary, readjust them.
-Check for fluoroscopy and exposure blocking when the tube unit support arm is not in the exposure position.
Component Panel:
-Switch power off to the system.
-Check all cables laid for good condition and secure connection. If necessary, replace any damaged cables.
-Retighten any cable connections.
-Retighten the mounting screws of all components.
-Check P/C boards for secure seating.
-Check
Electronics:
- Electronic adjustment parameters per section.3.28
Accessories:
-Check all accessories present at the customer’s location.
-Hook in the table extension, tilt the table and check the factory set safety distance of 12 cm; if necessary, adjust it.
-Check engagement of the leg support on the profile rail.
-Check for secure mounting of the head cushion mount.
-Check the patient table pad for good condition.
-Check the flush bowl for damage and check the function.
General:
-Reinstall the unit base cover panel.
-Reinstall the unit base cover panels.
-Reinstall the base cover panels.
-Reinstall the system carriage cover panels.
-Clean the unit.
-Clean surfaces with one of the following tested disinfecting agents:
Tego 103, Kosolin, Misty
-Switch power back on.
-Carry out a function check according to 4.2.