SSI America 7SL Calibration Procedure, General Guidelines, Calibration Parameters

Page 30
+0.005% output
+0.001% range
+5 mV
+0.02%
+0.0025 decade better than 0.1°C 0.1 mA AC rms
1 mV
10 m
better than 0.1°C 0.1 mA AC rms

CALIBRATION PROCEDURE

Calibration parameters are logically divided into groups of two parameters each – initial scale value and final scale value. A calibration check is provided after entering the values of each group. It is also possible to perform a calibration check without making an entry: press the FUNC button twice when “OFF” is displayed. The instru- ment goes directly to the group check.

Before beginning calibration, be sure the internal DIP switch V101 is open.

WARNING: Perform the calibration procedure according to J106 jumper positions as shown in Figure 1. Other- wise, the stored calibration values may be lost.

Input Type

 

J106

 

 

1-2

3-4

5-6

7-8

T/C, RTD,

 

 

 

 

0-60 mV, 12-60 mV

close

open

open

open

(standard configuration)

 

 

 

 

0-5 V, 1-5V

open

close

open

open

 

 

 

 

 

0-10V, 2-10 V

open

open

close

open

 

 

 

 

 

0-20 mA, 4-20 mA

open

open

open

close

 

 

 

 

 

Input CTTC R192=180K

 

 

 

SH120=Closed CH101=Open

 

 

 

Figure 1.

J106 Settings

 

 

General Guidelines

a)The instrument should be mounted in its case in order to keep the internal temperature constant.

b)Ambient temperature should be stable. Avoid drift due to air conditioning or other mechanical devices.

c)Relative humidity should not exceed 70%.

d)Minimum warm up time of at least 20 minutes.

e)Operate as much as possible in a noise free environment.

f)during calibration, connect one input at a time to the rear terminal block.

g)Use calibrators with the following:

Accuracy

TC Input:

RTD Input:

CJ Compensation:

Current Transformer:

Resolution

TC Input:

RTD Input:

CJ Compensation:

Current Transformer:

Calibration Parameters

Following is a complete list of calibration symbols: Code Parameter

tL

TC Input Initial Scale Value

tH

TC Input Final Scale Value

t.TC Input Check

rJ

Cold Junction Compensation

rJ.

Cold Junction Compensation Check

PL

RTD Input Initial Scale Value

PH

RTD Input Final Scale Value

P.RTD Input Check

AL

Current Input Initial Scale Value

AH

Current Input Final Scale Value

A.Current Input Check

nL

5

Volt Input Initial Scale Value

nH

5

Volt Input Final Scale Value

n.5 Volt Input Check

UL

10

Volt Input Initial Scale Value

UH

10

Volt Input Final Scale Value

U.10 Volt Input Check

3 0

7sl-6-00.p65

30

9/17/02, 2:44 PM

Image 30
Contents Model 7SL HIGH/LOW Limitrol 16 DIN, Four Digit HIGH/LOW Limitrol Congratulations Guide to simple set-upTable of Contents USE Wire Suitable For 75C MinimumMounting Requirements Dimensions and Panel Cutout Wiring Guidelines Terminal LayoutMeasuring Inputs TC Input RTD Linear InputRTD Input Logic Input7sl-1-00.p65 17/02, 239 PM Relay Outputs OUTSerial Interface CommonPower Line and grounding Power SupplyPreliminary Hardware Settings Limitrol with RS-485 Limitrol without RS-485Configuration KEY Functions Configuration ProcedureC2 = Acknowledgment Mode R5 = Offset AdjustmentR7 = Alarm Action on Fault R1 = Input Type and Range Value R2 = Decimal Point PositionControl Output Function C3 = Reset at Power-up C4 = Reset MemoryP2 = Alarm configuration = Safety Lock P3 = Alarm ActionAlarm Standby Function OFF =T1 = Timeout Selection tn10 = 10 second timeout Indicators Operating ModeNormal Display Mode ResetKey Functions in Normal Display Mode Parameter Display ModeKey Functions in Parameter Display Mode Operating Parameters Error List Error MessagesError Messages Dealing with Error MessagesDefault Operating Parameters List Default ParametersLoading Default Operating Parameters Loading Default Configuration ParametersDefault Configuration Parameter List ParaSpecifications Common Mode Rejection Ratio Display Update Time RangeTemperature Drift Normal Mode Rejection RatioAlarm Function Acknowledgment ModePower-Up Mode Type of AlarmNema ReferencesCEI 70-1 IEC ENV50140Calibration Procedure General GuidelinesCalibration Parameters TC Input Check ProcedureTH TC Input Final Scale Value Entering Calibration ValuesRJ. Cold Junction Compensation Check PH RTD Input Final Scale ValueRTD Check PL RTD Input Initial Scale ValueNH 5 Volt Input Final Scale Value AH Current Input Final Scale ValueCurrent Input Check Volt Input CheckUL 10 Volt Input Initial Scale Value UH 10 Volt Input Final Scale ValueMaintenance Manual Part No 7SL