SSI America 7SL instruction manual 7sl-1-00.p65 17/02, 239 PM

Page 9

Thermocouple Compensating Cable Color Codes.

Thermocouple

 

British

 

American

 

German

 

French

Material

 

BS 1843

ANSI MC 96.1

DIN 43710

NFE 18-001

T Copper Constantan

+

White

+

Blue

+

Red

+

Yellow

 

-

Blue

-

Red

-

Brown

-

Blue

 

 

Blue

 

Blue

 

Brown

 

Blue

J/L Iron Constantan

+

Yellow

+

White

+

Red

+

Yellow

 

-

Blue

-

Red

-

Blue

-

Black

 

 

Black

 

Black

 

Blue

 

Black

K Nickel Chromium

+

Brown

+

Yellow

+

Red

+

Yellow

Nickel Aluminum

-

Blue

-

Red

-

Green

-

Purple

 

 

Red

 

Yellow

 

Green

 

Yellow

R Platinum/Platinum

+

White

+

Black

+

Red

+

White

13% Rhodium

-

Blue

-

Red

-

White

-

Green

 

 

Green

 

Green

 

White

 

Green

S Platinum/Platinum

+

White

+

Black

+

Red

+

White

10% Rhodium

-

Blue

-

Red

-

White

-

Green

 

 

Green

 

Green

 

White

 

Green

E Chromel

+

Brown

+

Violet

 

 

Constantan

-

Blue

-

Red

 

 

 

 

Brown

 

Violet

 

 

 

 

B Platinum 30% Rh

 

+

Grey

 

 

Platinum 6% Rh

 

-

Red

 

 

 

 

 

 

Grey

 

 

N Nicrosil / Nisil

 

 

 

 

9

7sl-1-00.p65

9

9/17/02, 2:39 PM

Image 9
Contents Model 7SL HIGH/LOW Limitrol 16 DIN, Four Digit HIGH/LOW Limitrol Guide to simple set-up CongratulationsUSE Wire Suitable For 75C Minimum Table of ContentsMounting Requirements Dimensions and Panel Cutout Wiring Guidelines Terminal LayoutMeasuring Inputs TC Input RTD Input Linear InputRTD Logic Input7sl-1-00.p65 17/02, 239 PM OUT Relay OutputsCommon Serial InterfacePower Line and grounding Power SupplyPreliminary Hardware Settings Limitrol without RS-485 Limitrol with RS-485Configuration Procedure Configuration KEY FunctionsR7 = Alarm Action on Fault R5 = Offset AdjustmentC2 = Acknowledgment Mode R1 = Input Type and Range Value R2 = Decimal Point PositionControl Output Function C3 = Reset at Power-up C4 = Reset MemoryP2 = Alarm configuration Alarm Standby Function P3 = Alarm Action= Safety Lock OFF =T1 = Timeout Selection tn10 = 10 second timeout Normal Display Mode Operating ModeIndicators ResetKey Functions in Normal Display Mode Parameter Display ModeKey Functions in Parameter Display Mode Operating Parameters Error Messages Error MessagesError List Dealing with Error MessagesLoading Default Operating Parameters Default ParametersDefault Operating Parameters List Loading Default Configuration ParametersPara Default Configuration Parameter ListSpecifications Temperature Drift Display Update Time RangeCommon Mode Rejection Ratio Normal Mode Rejection RatioPower-Up Mode Acknowledgment ModeAlarm Function Type of AlarmCEI 70-1 IEC ReferencesNema ENV50140Calibration Procedure General GuidelinesCalibration Parameters TH TC Input Final Scale Value ProcedureTC Input Check Entering Calibration ValuesRTD Check PH RTD Input Final Scale ValueRJ. Cold Junction Compensation Check PL RTD Input Initial Scale ValueCurrent Input Check AH Current Input Final Scale ValueNH 5 Volt Input Final Scale Value Volt Input CheckUH 10 Volt Input Final Scale Value UL 10 Volt Input Initial Scale ValueMaintenance Manual Part No 7SL