SSI America 7SL UL 10 Volt Input Initial Scale Value, UH 10 Volt Input Final Scale Value

Page 34

UL 10 VOLT Input Initial Scale Value

a)Connect calibrator and instrument as shown in Figure 9.

10

9

Figure 9. Calibrator Connection

b)The upper display shows “OFF”, the lower displays shows “UL”.

c)Set calibrator to 0.000 V (even if the minimum range value is 2 V).

d)Push the button; the display changes to “ON”.

e)After a few seconds, start calibration by pushing the FUNC button. When this calibration is complete, the instrument will go to the next parameter.

UH 10 Volt Input Final Scale Value

a)Set the calibrator to 10.000 V.

b)Push the button; the displays will show “ON” and “10UH”.

c)After a few seconds, start calibration by pressing the FUNC button. When this calibration is complete, the instrument will go to the next parameter.

U.10 Volt Input Check

The display shows “10UH.” followed by a number show- ing the measured value in counts.

a)Check the calibration by setting:

0.000 V – the readout must be equal to “U.000 00” ± 10 counts;

10.000 V – the readout must be equal to “U.300 00” ± 10 counts;

5.000 V – the readout must be equal to “U.150 00” ± 10 counts.

b)Push the FUNC button.

3 4

7sl-6-00.p65

34

9/17/02, 2:44 PM

Image 34
Contents Model 7SL HIGH/LOW Limitrol 16 DIN, Four Digit HIGH/LOW Limitrol Congratulations Guide to simple set-upTable of Contents USE Wire Suitable For 75C MinimumMounting Requirements Dimensions and Panel Cutout Terminal Layout Wiring GuidelinesMeasuring Inputs TC Input RTD Linear InputRTD Input Logic Input7sl-1-00.p65 17/02, 239 PM Relay Outputs OUTSerial Interface CommonPower Supply Power Line and groundingPreliminary Hardware Settings Limitrol with RS-485 Limitrol without RS-485Configuration KEY Functions Configuration ProcedureC2 = Acknowledgment Mode R5 = Offset AdjustmentR7 = Alarm Action on Fault R1 = Input Type and Range Value R2 = Decimal Point PositionControl Output Function C4 = Reset Memory C3 = Reset at Power-upP2 = Alarm configuration = Safety Lock P3 = Alarm ActionAlarm Standby Function OFF =T1 = Timeout Selection tn10 = 10 second timeout Indicators Operating ModeNormal Display Mode ResetParameter Display Mode Key Functions in Normal Display ModeKey Functions in Parameter Display Mode Operating Parameters Error List Error MessagesError Messages Dealing with Error MessagesDefault Operating Parameters List Default ParametersLoading Default Operating Parameters Loading Default Configuration ParametersDefault Configuration Parameter List ParaSpecifications Common Mode Rejection Ratio Display Update Time RangeTemperature Drift Normal Mode Rejection RatioAlarm Function Acknowledgment ModePower-Up Mode Type of AlarmNema ReferencesCEI 70-1 IEC ENV50140General Guidelines Calibration ProcedureCalibration Parameters TC Input Check ProcedureTH TC Input Final Scale Value Entering Calibration ValuesRJ. Cold Junction Compensation Check PH RTD Input Final Scale ValueRTD Check PL RTD Input Initial Scale ValueNH 5 Volt Input Final Scale Value AH Current Input Final Scale ValueCurrent Input Check Volt Input CheckUL 10 Volt Input Initial Scale Value UH 10 Volt Input Final Scale ValueMaintenance Manual Part No 7SL