6.4D/A Adjustment
6.4.1Unipolar Analog Output
1.Set JP1 to select unipolar. Connect VDM (+) to CN3
2.Write the digital value 0 to DAC. Trim VR2 to obtain 0V reading in the DVM
3.Write the digital value 4095 to DAC. TrimVR1 to obtain 10V reading in the DVM.
6.4.2Bipolar Analog Output
1.Set JP1 to select bipolar. Connect DVM (+) to CN3
2.Write the digital value 2048 to DAC. Trim VR2 to obtain 0V reading in the DVM.
3.Write the digital value 4095 to DAC. Trim VR1 to obtain +10V reading in the DVM.
A calibration utility is supported in the software CD which is included in the product package. The detailed calibration procedures and description can be found in the utility. Users only need to run the software calibration utility and follow the procedures. You will get the accurate measure data.
In normal condition, the
Calibration ∙67