Chapter 2
Configuring the HP E1368A/69A/70A Modules
Using This Chapter
This chapter shows how to connect external wiring to the Microwave
Switch Modules connectors, and how to configure the module for operation.
∙ Setting the Address Switch | Page 18 |
∙ Selecting the Interrupt Priority | Page 19 |
∙ Installing Switches on the HP E1369A | Page 20 |
∙ Installing a Switch or Attenuator on the |
|
HP E1370A | Page 26 |
Warnings and Cautions
Warning SHOCK HAZARD. Only
Caution MAXIMUM POWER. The maximum power that may be applied to any SMA input connector is 1 W (CW). The maximum voltage that may be applied to the EXTERNAL B+ terminal is 42Vpk.
CONNECTING +5V/+12V. For the Microwave Switch, the mainframe backplane +5V is fused at 4A, and the +12V line at 4A. The total current drawn by all coaxial switches connected to the Microwave Switch module must not exceed the fuse rating of the supplies (mainframe and/or external) used.
STATIC ELECTRICITY. Static electricity is a major cause of component failure. To prevent damage to the electrical components in the Microwave Switch modules, observe
Chapter 2 | Configuring the HP E1368A/69A/70A Modules 17 |