HP E1368A manual Selecting Switch Power

Page 24

Selecting Switch Power

After the coaxial switches have been installed (refer to “ Installing Switches on the HP E1369A” earlier in this chapter), the correct operating voltage must be selected. See installed switch specifications.

The Microwave Switch modules are capable of providing +5V (set at factory) or +12V using the mainframe power supply, or external voltages of up to 42Vpk can be connected. Determine the required operating voltage and proceed as instructed below. Refer to Figure 2-5 as required.

Caution MAXIMUM CURRENT. The maximum current that the control circuit can accommodate is 1 amp per switch. Maximum current also depends on the output capacity of the mainframe or power supply used.

Internal Power Verify that the F3 fuse (P/N 2110-0712) is installed. Position jumpers as instructed:

For 5V, jumper JM2 installed, and jumper JM151 removed.

For 12V, jumper JM151 installed, and jumper JM2 removed.

External Power Verify that the F3 fuse (P/N 2110-0712) is removed (unsolder if necessary). Connect the positive lead to EXT B+ on TB1, and the negative lead to EXT GND on TB1. Connect leads to external power supply. Observe polarity.

Caution MAXIMUM VOLTAGE. The maximum voltage that may be applied to the EXTernal POWER terminal is 42Vpk.

Maximum wire size is No. 16 AWG. Wire ends should be stripped 6 mm (~ 0.25 in.) and tinned to prevent single strands from shorting adjacent terminals.

It is recommended that each channel wire be identified (color coded or marked) as the connection is not visible when the Microwave Switch module is installed.

Verify that wires make good connections on screw terminals.

24 Configuring the HP E1368A/69A/70A Modules

Chapter 2

Image 24
Contents Contents Open OPEN? Scan Scanmode SCANMODE? Cpon HP E1368A/69A/70A Modules User’s Manual Contents Certification Documentation History April According to ISO/IEC Guide 22 and ENPage Business Reply Mail Page Using This Chapter Microwave Switch DescriptionGetting Started with the HP E1368A/69A/70A Chapter Power Supply Circuit HP E1368A SwitchesProgramming the Microwave Switch Instrument DefinitionCard Numbers for Multiple-Module Switchboxes Example Reset the switchbox and close channel Initial OperationConfiguring the HP E1368A/69A/70A Modules Setting the Address Switch Logical Address SelectionSelecting the Interrupt Priority Interrupt Priority SelectionInstalling Switches on the HP E1369A Installing Switches on the ModuleChapter Configuring the HP E1368A/69A/70A Modules Microwave Switch Installation Connecting Switches External To the Module Selecting Switch Power Connecting Field Wiring Selecting Switch VoltageInstalling a Switch Or Attenuator on Module Microwave Switch/Attenuator Installation Shows the internal schematic of a single attenuator section Using the HP E1368A/69A/70A Modules Microwave Switch CommandsReset Conditions Switching Channels Example Single Channel SwitchingPower To connect C to Port 1, execute 33363K 33364K 33365K 33366K Example Multiple Channel Switching Using EXTernal Example Transfer Switch Using Two Port Switches Example Transfer Switch Using One Port Switch To route signal B to the output, execute Example 4 x 1 MultiplexerScanning Channels Selects the Trigger SourceExample Scanning Channels Output 722TRIG EXT …Example Using the Scan Complete Bit Example Multiple Channel Scanning110END Detecting Error Conditions Recalling and Saving StatesRepeat Synchronizing the Microwave Switch Synchronizing InstrumentsQuerying the Microwave Switch Command Types Common Command FormatScpi Command Format HP E1368A/69A/70A Command Reference Chapter Semicolon between the commands. For example Scpi Command Reference ABORtExample Stopping a Scan with ABORt For the number of scanning cycles default is ∙ Number of Scans Use only numeric values between 1Armcoun Example Select Module #2 in a Switchbox for Monitoring DISPlayMONitorCARD ∙ Selecting a Specific Module to be Monitored UseExample Enabling the Monitor Mode Example Querying the Monitor Mode∙ Selecting the Module to be Monitored Use ∙ Stopping Continuous Scans See the ABORt command INITiateExample Enabling Continuous Scans Example Query Continuous Scanning State Example Query Trig Out Port Enable State OUTPutExample Enabling Trig Out Port Example Closing Microwave Switch Channels ROUTe∙ Closing Channels To close Example Opening Microwave Switch Channels ∙ Opening Channels To openExample Scanning Using External Device Or any combinationChapter HP E1368A/69A/70A Command Reference STATus ∙ Setting Bit 7 of the Status Byte Register Example Query the Standard Operation Status Enable RegisterSTATOPER? SYSTem CDEScription?Comments Microwave Switch Module Model Number Example Reading the Error QueueExample Reading the Model Number of a Card #1 Module ∙ Executing the TRIGgerIMMediate Command a channel list Example Advancing Scan Using the TRIGger CommandTRIGger ∙ Hold or BUS Source Remains If selected, the TRIGgerSOURce∙ Trig Out Port Shared by Switchboxes See the OUTPut command Example Scanning Using External TriggersScanning Using Bus Triggers Example Query the Trigger SourceIeee 488.2 Common Commands HP Command Module User’s Manual, or the ANSI/IEEE StandardCommand Quick Reference Scpi Commands for Quick ReferenceHP E1368A/69A/70A Command Reference Chapter Characteristic Impedence 50Ω Insertion Loss Maximum Input Power 1 Watt averageConnectors Used P1 No. Slots Operating TemperatureHP E1368A/69A/70A Specifications Appendix a HP E1368A/69A/70A Registers Register DefinitionsAddressing the Registers Reading the RegistersID and Device Type Registers Status/Control Register Channel Enable Register Writing to the RegistersHP E1368A/69A/70A Registers Appendix B HP E1368A/69A/70A Error Messages Page Index HP E1368A/69A/70A Modules User’s Manual IndexHP E1368A/69A/70A Modules User’s Manual Index HP-IB HP E1368A/69A/70A Modules User’s Manual Index HP E1368A/69A/70A Modules User’s Manual Index HP E1368A/69A/70A Modules User’s Manual Index