Chapter 4
HP E1368A/69A/70A Command Reference
Using This Chapter
This chapter describes Standard Commands for Programmable Instruments (SCPI) commands and summarizes IEEE 488.2 Common (*) Commands applicable to the Microwave Switch. See the HP 75000 Series B Mainframe HP E1300A/E1301A User’s Manual or the HP 75000 Series C HP E1406 User’s Manual for additional information on SCPI and common commands. This chapter contains the following sections:
∙ Command Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Page 47
∙ SCPI Command Reference . . . . . . . . . . . . . . . . . . . . . . . . . . Page 50
∙ IEEE 488.2 Common Commands . . . . . . . . . . . . . . . . . . . . . Page 70
∙ Command Quick Reference . . . . . . . . . . . . . . . . . . . . . . . . . . Page 71
Command Types
Common Command Format
Commands are separated into two types: IEEE 488.2 Common Commands and SCPI Commands.
The IEEE 488.2 standard defines the Common Commands that perform functions like reset,
*RST | *ESR 32 | *STB? |
SCPI Command Format
SCPI commands perform functions like closing switches, making measurements, querying instrument states, or retrieving data. A subsystem command structure is a hierarchical structure that usually consists of a top level (or root) command, one or more lower level commands, and their parameters. The following example shows part of a typical subsystem:
[ROUTe:]
CLOSe <channel_list>
SCAN <channel_list>
:MODE?
[ROUTe:] is the root command, CLOSe and SCAN are second level commands with parameters, and :MODE? is a third level command.