:OPERation:ENABle STATus:OPERation:ENABle <number> enables the Standard Operation Status Register to set a bit in the Status Byte Register. For the Microwave Switch, when bit 8 in the Standard Operation Status Register is set to 1, bit 7 in the Status Byte Register is set to 1.
Parameters
Parameter Name | Parameter Type | Range of Values | Default Value |
|
|
|
|
number | numeric | 0 - 32767 | 0 |
|
|
|
|
|
|
|
|
Comments ∙ Summary Bit: If bit 8 is 1 in the Standard Operation Status Enable Register, a transition of Event Register bit 8 to 1 will cause the associated summary bit to be true.
∙Maximum Value Entered: The maximum decimal weighted value used in the Microwave Switch module is 256 (bit 8 set true).
∙Setting Bit 7 of the Status Byte Register:
STATus:OPERation:ENABle 256 sets bit 7 of the Status Byte Register to 1 after bit 8 of the Standard Operation Status Enable Register is set to 1.
∙Related Commands: [ROUTe:]SCAN
Example Enable the Status Byte Register Bit 7
STAT:OPER:ENAB 256 | Enables bit 7 of the Standard |
| Operation Status Enable Register, |
| which enables bit 7 of Status Byte |
| Register. |
:OPERation:ENABle? STATus:OPERation:ENABle? returns the bit value of the Standard
Operation Status Enable Register.
Comments ∙ Output Format: Returns a decimal weighted value from 0 to 65535 indicating which bits are set true.
∙Maximum Value Returned: The maximum decimal weighted value used in the Microwave Switch module is 256 (bit 8 set true).
Example Query the Standard Operation Status Enable Register
STAT:OPER:ENAB? | Queries the Standard Operation |
| Status Enable Register. |
Chapter 4 | HP E1368A/69A/70A Command Reference 63 |