HP E1368A manual Scanning Using Bus Triggers, Example Query the Trigger Source

Page 69

Scanning Using Bus Triggers

This example uses bus triggering (TRIG:SOUR BUS) to scan channels 00 through 02 of a single-module switchbox. The trigger source to advance the scan is the *TRG command (as set with TRIGger:SOURce BUS). When INIT is executed, the scan is started and channel 00 is closed. Then, each *TRG command advances the scan to the next channel.

TRIG:SOUR BUS

Select interface (BUS) triggering.

SCAN (@100:102)

Scan channels 00 through 02.

INIT

Start scan, close channel 00.

*TRG

Advances scan using BUS

 

triggering.

increment loop

Increment loop count.

:SOURce? TRIGger:SOURce? returns the current trigger source for the switchbox. Command returns either BUS, EXT, HOLD, or IMM for trigger sources BUS, EXTernal, HOLD, or IMMediate, respectively.

Example Query the Trigger Source

TRIG:SOUR EXT

Sets trigger source to external.

TRIG:SOUR?

Query trigger source.

Chapter 4

HP E1368A/69A/70A Command Reference 69

Image 69
Contents Contents Open OPEN? Scan Scanmode SCANMODE? Cpon HP E1368A/69A/70A Modules User’s Manual Contents Certification Documentation History According to ISO/IEC Guide 22 and EN AprilPage Business Reply Mail Page Microwave Switch Description Using This ChapterGetting Started with the HP E1368A/69A/70A Chapter HP E1368A Switches Power Supply CircuitInstrument Definition Programming the Microwave SwitchCard Numbers for Multiple-Module Switchboxes Initial Operation Example Reset the switchbox and close channelConfiguring the HP E1368A/69A/70A Modules Logical Address Selection Setting the Address SwitchInterrupt Priority Selection Selecting the Interrupt PriorityInstalling Switches on the Module Installing Switches on the HP E1369AChapter Configuring the HP E1368A/69A/70A Modules Microwave Switch Installation Connecting Switches External To the Module Selecting Switch Power Selecting Switch Voltage Connecting Field WiringInstalling a Switch Or Attenuator on Module Microwave Switch/Attenuator Installation Shows the internal schematic of a single attenuator section Microwave Switch Commands Using the HP E1368A/69A/70A ModulesReset Conditions Example Single Channel Switching Switching ChannelsPower To connect C to Port 1, execute 33363K 33364K 33365K 33366K Example Multiple Channel Switching Using EXTernal Example Transfer Switch Using Two Port Switches Example Transfer Switch Using One Port Switch Example 4 x 1 Multiplexer To route signal B to the output, executeSelects the Trigger Source Scanning ChannelsOutput 722TRIG EXT … Example Scanning ChannelsExample Multiple Channel Scanning Example Using the Scan Complete Bit110END Recalling and Saving States Detecting Error ConditionsRepeat Synchronizing Instruments Synchronizing the Microwave SwitchQuerying the Microwave Switch Command Types Common Command FormatScpi Command Format HP E1368A/69A/70A Command Reference Chapter Semicolon between the commands. For example Scpi Command Reference ABORtExample Stopping a Scan with ABORt ∙ Number of Scans Use only numeric values between 1 For the number of scanning cycles default isArmcoun DISPlay Example Select Module #2 in a Switchbox for MonitoringMONitorCARD ∙ Selecting a Specific Module to be Monitored UseExample Enabling the Monitor Mode Example Querying the Monitor Mode∙ Selecting the Module to be Monitored Use ∙ Stopping Continuous Scans See the ABORt command INITiateExample Enabling Continuous Scans Example Query Continuous Scanning State Example Query Trig Out Port Enable State OUTPutExample Enabling Trig Out Port Example Closing Microwave Switch Channels ROUTe∙ Closing Channels To close ∙ Opening Channels To open Example Opening Microwave Switch ChannelsOr any combination Example Scanning Using External DeviceChapter HP E1368A/69A/70A Command Reference STATus Example Query the Standard Operation Status Enable Register ∙ Setting Bit 7 of the Status Byte RegisterSTATOPER? CDEScription? SYSTemComments Microwave Switch Module Model Number Example Reading the Error QueueExample Reading the Model Number of a Card #1 Module Example Advancing Scan Using the TRIGger Command ∙ Executing the TRIGgerIMMediate Command a channel listTRIGger ∙ Hold or BUS Source Remains If selected, the TRIGgerSOURceExample Scanning Using External Triggers ∙ Trig Out Port Shared by Switchboxes See the OUTPut commandExample Query the Trigger Source Scanning Using Bus TriggersHP Command Module User’s Manual, or the ANSI/IEEE Standard Ieee 488.2 Common CommandsScpi Commands for Quick Reference Command Quick ReferenceHP E1368A/69A/70A Command Reference Chapter Maximum Input Power 1 Watt average Characteristic Impedence 50Ω Insertion LossConnectors Used P1 No. Slots Operating TemperatureHP E1368A/69A/70A Specifications Appendix a Register Definitions HP E1368A/69A/70A RegistersAddressing the Registers Reading the RegistersID and Device Type Registers Writing to the Registers Status/Control Register Channel Enable RegisterHP E1368A/69A/70A Registers Appendix B HP E1368A/69A/70A Error Messages Page HP E1368A/69A/70A Modules User’s Manual Index IndexHP E1368A/69A/70A Modules User’s Manual Index HP-IB HP E1368A/69A/70A Modules User’s Manual Index HP E1368A/69A/70A Modules User’s Manual Index HP E1368A/69A/70A Modules User’s Manual Index