Connecting Field Wiring
Figure 2-8 shows the internal switch diagram for the various HP single-pole, multi-throw switches. All switches are shown in the "OPEN" position. To minimize loss at high frequencies, use the following guidelines when making connections.
∙For frequencies up to 18 GHz, use a good quality flexible type cable and SMA connectors.
∙For frequencies >18 GHz, use semi-rigid type cable and APC 3.5 connectors.
Figure 2-8. Internal Switch Diagrams
Figure 2-9 shows the internal schematic of a single attenuator section.
Figure 2-9. Attenuator
28 Configuring the HP E1368A/69A/70A Modules | Chapter 2 |