Bacharach H25-IR Selecting Internal or External Calibration, Internal Calibration Selection

Page 25

H25-IR

Calibration

 

 

3 Calibration

When the instrument is supplied with an internal leak source, the operator can choose between two types of calibration: Internal or External. Internal calibration uses the instrument’s built-in leak source, while External calibration uses a customer supplied calibrated leak source.

If the instrument is supplied without a leak source, then the option of selecting the calibration type is removed from the menu system. The only option for this model of instrument is to calibrate using an external leak source.

3.1 Selecting Internal or External Calibration

Use the Main Menu > CAL DATA > CAL TYPE function to select either Internal or External calibration. If External calibration is selected, then also use the EXT RATE and EXT GAS TYPE functions to enter the characteristics of the external leak source.

Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to highlight “CAL DATA”, and then press ENT to display the Calibration Data menu.

Main Menu:

SYSTEM

 

 

 

LOG

 

 

 

SETUPS

Calibration Data Menu:

CAL DATA

DIAGNOSTICS

 

CAL TYPE

 

 

 

EXT RATE

 

 

EXT GAS TYPE

 

 

CAL FACTOR

 

 

RESTORE FACTORY CAL

3.1.1 Internal Calibration Selection

From the Calibration Data menu, use the Up and Down buttons to highlight “CAL TYPE”, and then press ENT to display its setup screen. Use the Up and Down buttons to highlight “INTERNAL”,

and press ENT to make the selection and return to the Calibration Data menu. Press ESC as required to return to the Data Display screen.

CAL TYPE

INTERNAL

 

EXTERNAL

Instruction 3015-4342

21

Image 25
Contents Infrared Refrigerant Gas Leak Detector H25-IRRegister Your Warranty by Visiting WarrantyTable of Contents Parts and Service Hazard Symbols Safety PrecautionsAC Power and Grounding IntroductionExplosive Atmosphere CleaningWorking Inside the Instrument Misuse and Modifications to Instrument Case of MalfunctionUnpacking and Initial Checks Ordering InformationFeatures and Capabilities Leak Rate = Flow Rate x Gas Concentration Functional OverviewFront View Names and Functions of ComponentsRear View Interior Standard Probe with LED and Programmable Pushbutton Standard ProbeSmart Probe Smart Probe Keylock Function System Menu Menu SystemConnecting the AC Power Cord Connecting the ProbePreparing for Operation Turning on the InstrumentMain Menu System Setup Menu System SetupGas Type Leak Units4 4-20 mA Loop RelaysReset Defaults Pchk ModeBrightness ClockSetup Name Configuring the Setups FunctionRange Setpoint Feedback ModeSearch Auto Zero CalibrationExternal Relay Connections Typical Relay 1 Wiring4-40 mA Current Loop Connection MA Current Loop ConnectorPreparing for Operation H25-IR Selecting Internal or External Calibration CalibrationInternal Calibration Selection Calibration Data MenuExternal Calibration Selection and Setup Internal Calibration CALExternal Calibration CAL Factor ProcedureMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Restoring Factory Calibration Adjusting the CAL FactorCalibration H25-IR Power Using the H25-IRInitial Setup OperationRange Selection Setup SelectionSearching for a Leak Volume AdjustmentMeasuring the PPM Level Measuring the Leak RateWorking with Logged Measurements Logging Measured ReadingsLog Menu Clearing the Log What is PCHK? What to Do when a Fault OccursDiagnostics Menu RS-232 CommunicationOperation H25-IR Diagnostics Menu MaintenanceSensor Data Last Fault IR Emitter Factory MenuPressure Factory Factory MenuLeak Rate Auto GainLeak Gas Leak TempSensor Cal Leak TempcoProbe Type Model SelectFirmware Items Required Flashing MemorySelect Setup Communications Select File Load Flash Image Removing the Top Cover Part ReplacementProbe Tip Filter & O-Ring Replacement Fuse ReplacementInternal Filter Replacement Replacement Parts Parts and ServiceUnited States Service CentersCFCs, HCFCs, HFCs and Halogens SpecificationsSpecifications H25-IR Continuous Appendix a Factory DefaultsAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA