Bacharach H25-IR manual External Calibration Selection and Setup

Page 26

Calibration

H25-IR

 

 

3.1.2 External Calibration Selection and Setup

When using an external leak source, the instrument must be setup for external calibration along with entering the leak rate and gas type of the external leak source.

Select external calibration and enter the parameters of the external leak source as follows:

Select External Calibration: From the Calibration Data menu, use the Up and Down buttons to highlight “CAL TYPE”, and then press ENT to display its setup screen. Use the Up and Down buttons to highlight “EXTERNAL”, and press ENT to make the selection and return to the Calibration Data menu.

CAL TYPE

INTERNAL

 

EXTERNAL

Enter External Leak Rate: From the Calibration Data menu, highlight “EXT RATE” and press ENT to display its setup screen. Use the Left and Right buttons to move the blinking cursor across the screen to highlight the digit to be modified, and then use the Up and Down buttons to modify that digit. Press ENT to save the new external leak rate value and return to the Calibration Data menu.

IMPORTANT! Be sure to enter the leak rate using the same units of measure as displayed on the EXT RATE screen. In this example, the units of measure is g/yr. If your leak source is marked in Oz/yr, then multiply that value by a factor of 28.35. For example: If the leak source is marked 0.9 Oz/yr, then enter a leak rate of (0.9 x 28.35=) 25.5 g/yr.

NOTE: The minimum calibration leak rate is 2 g/yr.

EXT RATE 25.5 g/yr

Enter External Gas Type: From the Calibration Data menu, highlight “EXT GAS TYPE” and press ENT to display its setup screen. Use the Up and Down buttons to highlight the type of gas contained in the external leak source. Press ENT to save the new external gas type and return to the Calibration Data menu.

NOTE: The gases displayed depend on which optical bench is installed. Refer to Section “7 Specifications.”

EXT GAS TYPE < R134A R401A R402A < R402B

Press ESC as required to return to the Data Display screen.

22

Instruction 3015-4342

Image 26
Contents H25-IR Infrared Refrigerant Gas Leak DetectorWarranty Register Your Warranty by VisitingTable of Contents Parts and Service AC Power and Grounding Safety PrecautionsHazard Symbols IntroductionWorking Inside the Instrument CleaningExplosive Atmosphere Misuse and Modifications to Instrument Case of MalfunctionOrdering Information Unpacking and Initial ChecksFeatures and Capabilities Functional Overview Leak Rate = Flow Rate x Gas ConcentrationNames and Functions of Components Front ViewRear View Interior Standard Probe Standard Probe with LED and Programmable PushbuttonSmart Probe Smart Probe Keylock Function Menu System System MenuPreparing for Operation Connecting the ProbeConnecting the AC Power Cord Turning on the InstrumentGas Type System SetupMain Menu System Setup Menu Leak UnitsRelays 4 4-20 mA LoopBrightness Pchk ModeReset Defaults ClockRange Configuring the Setups FunctionSetup Name Feedback Mode SetpointCalibration Search Auto ZeroTypical Relay 1 Wiring External Relay ConnectionsMA Current Loop Connector 4-40 mA Current Loop ConnectionPreparing for Operation H25-IR Internal Calibration Selection CalibrationSelecting Internal or External Calibration Calibration Data MenuExternal Calibration Selection and Setup CAL Internal CalibrationExternal Calibration Material Required ProcedureCAL Factor Improving PPM Measurement Accuracy using the CAL Factor Adjusting the CAL Factor Restoring Factory CalibrationCalibration H25-IR Initial Setup Using the H25-IRPower OperationSetup Selection Range SelectionVolume Adjustment Searching for a LeakMeasuring the Leak Rate Measuring the PPM LevelLog Menu Logging Measured ReadingsWorking with Logged Measurements Clearing the Log Diagnostics Menu What to Do when a Fault OccursWhat is PCHK? RS-232 CommunicationOperation H25-IR Sensor Data MaintenanceDiagnostics Menu Last Fault Pressure Factory MenuIR Emitter Factory Factory MenuLeak Gas Auto GainLeak Rate Leak TempLeak Tempco Sensor CalFirmware Model SelectProbe Type Flashing Memory Items RequiredSelect Setup Communications Select File Load Flash Image Part Replacement Removing the Top CoverFuse Replacement Probe Tip Filter & O-Ring ReplacementInternal Filter Replacement Parts and Service Replacement PartsService Centers United StatesSpecifications CFCs, HCFCs, HFCs and HalogensSpecifications H25-IR Appendix a Factory Defaults ContinuousAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA