Calibration | |
|
|
3.3 External Calibration
External calibration uses a customer supplied leak source that matches the gas to be monitored. Properties of the leak source must be entered by the operator using the EXT RATE and EXT GAS TYPE functions as described in Section 3.1.2 External Calibration Selection and Setup. The EXT RATE entry is the number displayed as the “target” value during calibration.
IMPORTANT! The operator is responsible for ensuring that the external
1.Turn ON instrument and allow it to complete its 120 second
2.Press the front panel CAL button to display the Calibration screen. Note that the display must be reading a leak rate of “0.00” when the CAL button is pressed; otherwise, the message “LEAK RATE MUST BE ZERO TO CALIBRATE” will appear. If necessary, press the ZERO button to zero the gas display before pressing CAL.
0.89TARGET Oz/yr
O/YR R134A
R134A
0.00MEASURED SETUP
O/YR EXT CAL
3.Ensure that the instrument is set up for “EXT CAL” (external calibration) as indicated at the
4.Insert probe tip into the Calibration Port of the external leak source.
5.The top number of the Calibration screen shows the “target” value (leak source rate), while the bottom number shows the current “measured” value. Keep the probe tip inserted in the Calibration Port for the duration of the calibration process, during which time the instrument performs a calibration at both a high and low flow rate.
0.89
0.89
TARGET | Oz/yr | |
OZ/YR | R134A | |
R134A | ||
SETUP | ||
MEASURED | ||
OZ/YR | EXT CAL |
6.After the calibration cycle is complete, the new calibration data is saved and the display automatically switches back to the Data Display screen. Note, however, if the PCHK Mode is set to AUTO (refer to Section 2.4.7 PCHK Mode), then a pressure check is performed prior to the display of the Data Display screen.
Oz/yr
R134A
SETUP
SM SRCH
NOTE: After calibration the instrument is automatically placed into its Search mode.
24 | Instruction |