Bacharach H25-IR manual Select Setup Communications

Page 48

Maintenance

H25-IR

 

 

7.From the C:\RFU folder, run the flash memory program by double clicking the file rfu.exe.

8.Select Setup > Communications.

9.Under “Serial Options”, select Use Serial Connection. Then set the Baud Rate to 115200, and select Enable Processor Detection.

Note that COM1 is the default Comm Port. If necessary, change the Comm Port to match the port that the instrument was connected to in Step 2.

Click OK.

10.Select Setup > File Locations and verify that the coldload.bin, pilot.bin, and flash.ini files are all located in the C:\RFU folder.

Click OK if these files are in the correct folder. If necessary, use the browse buttons in the Choose File Locations dialog box to locate each of these files.

44

Instruction 3015-4342

Image 48
Contents H25-IR Infrared Refrigerant Gas Leak DetectorWarranty Register Your Warranty by VisitingTable of Contents Parts and Service Safety Precautions Hazard SymbolsAC Power and Grounding IntroductionCleaning Explosive AtmosphereWorking Inside the Instrument Misuse and Modifications to Instrument Case of MalfunctionOrdering Information Unpacking and Initial ChecksFeatures and Capabilities Functional Overview Leak Rate = Flow Rate x Gas ConcentrationNames and Functions of Components Front ViewRear View Interior Standard Probe Standard Probe with LED and Programmable PushbuttonSmart Probe Smart Probe Keylock Function Menu System System MenuConnecting the Probe Connecting the AC Power CordPreparing for Operation Turning on the InstrumentSystem Setup Main Menu System Setup MenuGas Type Leak UnitsRelays 4 4-20 mA LoopPchk Mode Reset DefaultsBrightness ClockConfiguring the Setups Function Setup NameRange Feedback Mode SetpointCalibration Search Auto ZeroTypical Relay 1 Wiring External Relay ConnectionsMA Current Loop Connector 4-40 mA Current Loop ConnectionPreparing for Operation H25-IR Calibration Selecting Internal or External CalibrationInternal Calibration Selection Calibration Data MenuExternal Calibration Selection and Setup CAL Internal CalibrationExternal Calibration Procedure CAL FactorMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Adjusting the CAL Factor Restoring Factory CalibrationCalibration H25-IR Using the H25-IR PowerInitial Setup OperationSetup Selection Range SelectionVolume Adjustment Searching for a LeakMeasuring the Leak Rate Measuring the PPM LevelLogging Measured Readings Working with Logged MeasurementsLog Menu Clearing the Log What to Do when a Fault Occurs What is PCHK?Diagnostics Menu RS-232 CommunicationOperation H25-IR Maintenance Diagnostics MenuSensor Data Last Fault Factory Menu IR EmitterPressure Factory Factory MenuAuto Gain Leak RateLeak Gas Leak TempLeak Tempco Sensor CalModel Select Probe TypeFirmware Flashing Memory Items RequiredSelect Setup Communications Select File Load Flash Image Part Replacement Removing the Top CoverFuse Replacement Probe Tip Filter & O-Ring ReplacementInternal Filter Replacement Parts and Service Replacement PartsService Centers United StatesSpecifications CFCs, HCFCs, HFCs and HalogensSpecifications H25-IR Appendix a Factory Defaults ContinuousAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA