Bacharach H25-IR manual Specifications, CFCs, HCFCs, HFCs and Halogens

Page 55

H25-IR

Specifications

 

 

7 Specifications

Gases Detected

Detects and measures, depending on instrument’s part number:

 

• CFCs, HCFCs, HFCs and Halogens

 

R11, R12, R22, R23, R113, R114, R123, R124, R134a, R227, R236fa,

 

R401A, R402A, R402A, R404A, R407A, R407C, R408A, R409A, R410A,

 

R500, R502, R503, R507, R508B, H1211, H1301, H2402, N1230, FA188

 

• R600a

 

• SF6

 

• CO2

 

• R227

Detection Method

Infrared

Measurement Units

oz/yr, g/yr, mL/s, PaM3/s, ppm

Measurement Range

0.03 to 5.00 oz/yr

 

0.85 to 142 g/yr

 

0.08 x 10–4to 100 x 10–4mL/s

 

0.08 x 10–5to 100 x 10–5PaM3/s

 

0 to 1,000 ppm

Operating Modes

Search: detects presence and location of gas leaks

 

Measure: calculates and displays leak rate

Response Time

0.5 to 1 second (dependent on probe-hose length)

Sensitivity

0.03 oz/yr (0.9 g/yr) all gases

Resolution

0.01 oz/yr (0.1 g/yr)

Accuracy:

 

Leak Rate

±1% of the accuracy value marked on tag attached to leak source

PPM

±10% of reading (can be improved to ±1% if recalibrated using a known

 

concentration of refrigerant gas)

Temperature Drift

±0.8% of reading per °C

Calibration

Internal: calibration performed using the unit’s built-in leak source

 

External: calibration performed using a customer supplied leak

 

source of a known gas type and leak rate

Relays

Four SPDT relays (rated 3 A, 240 VAC), each programmable to energize

 

under 11 different operating conditions

Audio Speaker

Audible indication of leak level

AC Power

100 to 240 VAC, 50/60 Hz

Power Consumption

15 Watts

Warm-Up Time

2 minutes (30 minutes to meet published specifications)

Operating Temperature

32 to 122 °F (0 to 50 °C)

Storage Temperature

4 to 122 °F (20 to 50 °C)

Ambient Humidity

5 to 90% RH, non-condensing

Dimensions

4 x 10.75 x 15.5 in.

 

(101.6 x 273.1 x 393.7 mm)

Weight

18 lb (8.2 kg)

Instruction 3015-4342

51

Image 55
Contents Infrared Refrigerant Gas Leak Detector H25-IRRegister Your Warranty by Visiting WarrantyTable of Contents Parts and Service Introduction Safety PrecautionsHazard Symbols AC Power and GroundingMisuse and Modifications to Instrument Case of Malfunction CleaningExplosive Atmosphere Working Inside the InstrumentUnpacking and Initial Checks Ordering InformationFeatures and Capabilities Leak Rate = Flow Rate x Gas Concentration Functional OverviewFront View Names and Functions of ComponentsRear View Interior Standard Probe with LED and Programmable Pushbutton Standard ProbeSmart Probe Smart Probe Keylock Function System Menu Menu SystemTurning on the Instrument Connecting the ProbeConnecting the AC Power Cord Preparing for OperationLeak Units System SetupMain Menu System Setup Menu Gas Type4 4-20 mA Loop RelaysClock Pchk ModeReset Defaults BrightnessSetup Name Configuring the Setups FunctionRange Setpoint Feedback ModeSearch Auto Zero CalibrationExternal Relay Connections Typical Relay 1 Wiring4-40 mA Current Loop Connection MA Current Loop ConnectorPreparing for Operation H25-IR Calibration Data Menu CalibrationSelecting Internal or External Calibration Internal Calibration SelectionExternal Calibration Selection and Setup Internal Calibration CALExternal Calibration CAL Factor ProcedureMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Restoring Factory Calibration Adjusting the CAL FactorCalibration H25-IR Operation Using the H25-IRPower Initial SetupRange Selection Setup SelectionSearching for a Leak Volume AdjustmentMeasuring the PPM Level Measuring the Leak RateWorking with Logged Measurements Logging Measured ReadingsLog Menu Clearing the Log RS-232 Communication What to Do when a Fault OccursWhat is PCHK? Diagnostics MenuOperation H25-IR Diagnostics Menu MaintenanceSensor Data Last Fault Factory Factory Menu Factory MenuIR Emitter PressureLeak Temp Auto GainLeak Rate Leak GasSensor Cal Leak TempcoProbe Type Model SelectFirmware Items Required Flashing MemorySelect Setup Communications Select File Load Flash Image Removing the Top Cover Part ReplacementProbe Tip Filter & O-Ring Replacement Fuse ReplacementInternal Filter Replacement Replacement Parts Parts and ServiceUnited States Service CentersCFCs, HCFCs, HFCs and Halogens SpecificationsSpecifications H25-IR Continuous Appendix a Factory DefaultsAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA