H25-IR

Maintenance

 

 

5.2.6 Leak TEMPCO

LEAK TEMPCO

0.54 %/degC

The Leak Temperature Coefficient value is set to match the temperature coefficient of the internal leak source as indicated on the tag attached to the leak source. Use the Left and Right buttons to move the blinking cursor across the screen to highlight the digit to be modified, and then use the Up and Down buttons to modify that digit. Press ENT to save the new Leak Temperature Coefficient value.

TIP: To convert a temperature coefficient from %/degF to % degC, multiply the

degF value by a factor of 1.8. For example: 0.3%/degF x 1.8 = 0.54%/degC.

5.2.7 Sensor Cal

PRESSURE

FLOW RATE

BOX TEMP

The Sensor Calibration function allows the service technician to manually calibrate the instrument’s Pressure, Flow Rate, and Box Temperature sensors. Use the Up and Down buttons to highlight the sensor to be calibrated, and then press ENT to display that sensor’s calibration screen.

5.2.7.1 Pressure Sensor Calibration Screen

ENTER AMBIENT PSIA 14.18

CURRENT OFFSET=0.00

Calibrate the pressure sensor by first measuring the ambient atmospheric pressure using an appropriate pressure measurement device calibrated to read PSIA. Next, adjust the value displayed on the Pressure Sensor Calibration screen to match the measured pressure reading by using the Left and Right buttons to move the blinking cursor across the screen to highlight the digit to be modified, and then using the Up and Down buttons to modify that digit. Press ENT to save the new ambient pressure value.

Tip: To convert inHg into PSIA, multiply inHg by 0.491. For example: 30.00 inHg x 0.491 = 14.73 PSIA.

5.2.7.2 Flow Rate Sensor Calibration Screen (Factory Calibration)

S-Mpm/dP

M-Mpm/dP

INT 0.0177/-1.33

0.0100/-2.28

FAC 0.0177/-1.33

0.0100/-2.28

ENTER TO SAVE INT AS FACTORY

Set the default factory calibration flow and pressure drop values by first performing an internal calibration as described in Section 3.2 Internal Calibration.

INT: Flow rate in moles/minute for both the Search and Measure modes after performing a calibration using the internal leak source, along with showing the manifold’s pressure drop.

FAC: Flow rate and pressure drop values that are currently stored as the default factory calibration values. These values are used to validate calibrations performed by the operator. A calibration error will occur if the operator’s calibration and factory values differ by more than 20%.

Press ENT to store the internal calibration numbers as factory values. These are the calibration values that are reset when RESTORE FACTORY CAL is selected from the CAL Data menu (refer to Section 3.5 Restoring Factory Calibration).

Instruction 3015-4342

41

Image 45
Contents Infrared Refrigerant Gas Leak Detector H25-IRRegister Your Warranty by Visiting WarrantyTable of Contents Parts and Service Hazard Symbols Safety PrecautionsAC Power and Grounding IntroductionExplosive Atmosphere CleaningWorking Inside the Instrument Misuse and Modifications to Instrument Case of MalfunctionUnpacking and Initial Checks Ordering InformationFeatures and Capabilities Leak Rate = Flow Rate x Gas Concentration Functional OverviewFront View Names and Functions of ComponentsRear View Interior Standard Probe with LED and Programmable Pushbutton Standard ProbeSmart Probe Smart Probe Keylock Function System Menu Menu SystemConnecting the AC Power Cord Connecting the ProbePreparing for Operation Turning on the InstrumentMain Menu System Setup Menu System SetupGas Type Leak Units4 4-20 mA Loop RelaysReset Defaults Pchk ModeBrightness ClockConfiguring the Setups Function Setup NameRange Setpoint Feedback ModeSearch Auto Zero CalibrationExternal Relay Connections Typical Relay 1 Wiring4-40 mA Current Loop Connection MA Current Loop ConnectorPreparing for Operation H25-IR Selecting Internal or External Calibration CalibrationInternal Calibration Selection Calibration Data MenuExternal Calibration Selection and Setup Internal Calibration CALExternal Calibration Procedure CAL FactorMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Restoring Factory Calibration Adjusting the CAL FactorCalibration H25-IR Power Using the H25-IRInitial Setup OperationRange Selection Setup SelectionSearching for a Leak Volume AdjustmentMeasuring the PPM Level Measuring the Leak RateLogging Measured Readings Working with Logged MeasurementsLog Menu Clearing the Log What is PCHK? What to Do when a Fault OccursDiagnostics Menu RS-232 CommunicationOperation H25-IR Maintenance Diagnostics MenuSensor Data Last Fault IR Emitter Factory MenuPressure Factory Factory MenuLeak Rate Auto GainLeak Gas Leak TempSensor Cal Leak TempcoModel Select Probe TypeFirmware Items Required Flashing MemorySelect Setup Communications Select File Load Flash Image Removing the Top Cover Part ReplacementProbe Tip Filter & O-Ring Replacement Fuse ReplacementInternal Filter Replacement Replacement Parts Parts and ServiceUnited States Service CentersCFCs, HCFCs, HFCs and Halogens SpecificationsSpecifications H25-IR Continuous Appendix a Factory DefaultsAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA