Operation | |
|
|
4.3 What to Do when a Fault Occurs
When the instrument’s diagnostics detects that a fault has occurred, the word “FAULT” appears in place of the Gas Type, and an audible beep occurs every 2 seconds.
Oz/yr
FAULT
SETUP SM SRCH
Fault has occurred
The cause of the fault can be determined by viewing the Diagnostic menu’s Last Fault function. To display the Diagnostics menu, first press both the ENT and ESC buttons on either the instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to highlight “DIAGNOSTICS” and then press ENT to display the Diagnostics menu.
Main Menu:
SYSTEM LOG SETUPS CAL DATA DIAGNOSTICS FACTORY
Diagnostics Menu:
SENSOR DATA
LAST FAULT
IR EMITTER
PRESSURE
Select “LAST FAULT” to display the Last Fault screen. Refer to Section 5.1.2 Last Fault for a listing of the type of faults that are tracked by the diagnostics, and for information on how to possibly correct the fault.
Once the cause of the fault has been corrected, the FAULT message and the audible beep are automatically turned OFF.
4.4 What is PCHK?
The instrument maintains its accuracy under varying atmospheric conditions by performing an atmospheric pressure check either once every hour, or immediately after calibration (refer to Section Error! Reference source not found. Error! Reference source not found.). When a PCHK is performed, the unit beeps once, the pump stops, a valve opens exposing the pressure sensor to the atmosphere, and “PCHK” appears at the
Oz/yr
R134A
SETUP
PCHK
Pressure check being performed
4.5 RS-232 Communication
The
Instruction | 35 |