Calibration | |
|
|
7.Compute the new CAL Factor by dividing the actual leak rate of the external leak source by the reading recorded in Step 6. For example, if the leak rate of the external leak source is 0.4 oz/yr and the measured rate was 0.43 oz/yr, then the new CAL Factor is 0.4 ⎟ 0.43 = 0.930.
This new CAL Factor establishes the relationship between the internal calibration gas and the gas type selected to be measured.
8.Enter the new CAL Factor as described in Section 3.4.3 Adjusting the CAL Factor. Note that the displayed Gas Type should match the gas selected in Step 2.
9.From this point on the operator can perform an internal calibration and get accurate measurements for other gas types using whatever gas is contained in the internal leak source.
3.4.2 Improving PPM Measurement Accuracy using the CAL Factor
The CAL Factor can be adjusted to improve the accuracy of the ppm measurement by calibrating the instrument on the gas being measured. The accuracy of the ppm measurement after performing this procedure will equal the accuracy of the gas sample, ±1%. Note that the accuracy of the instrument’s ppm measurement depends on the accuracy of the gas sample.
The new CAL Factor is determined by sampling a known dilution of gas whose accuracy is better than the ppm measurement accuracy as stated in Section 7 Specifications. The concentration of the sample must also be corrected for ambient temperature and pressure at the time of calibration. A minimum sample size of 5 liters is required.
Calibration is best performed at or near full scale, it can however, be done at any concentration and ideally in the range where maximum accuracy is desired down to, but not below, 100 ppm.
The instrument should be operating for at least 30 minutes prior to performing a calibration.
Verify that the current CAL Factor is set to 1.000 by selecting the Main Menu > CAL DATA > CAL FACTOR function. If necessary, reset the CAL Factor to 1.000. Refer to Section 3.4.3 Adjusting the CAL Factor.
With the instrument operating in its ppm measurement mode (refer to Section 4.2.8 Measuring the PPM Level), insert the probe tip into the
IMPORTANT! If the instrument is switched back to measuring the
26 | Instruction |