Bacharach H25-IR manual Internal Calibration

Page 27

H25-IR

Calibration

 

 

3.2 Internal Calibration

Internal calibration uses the instrument’s built-in leak source. Properties of the leak source (i.e. gas type, leak rate, temperature coefficient, and specification temperature) are entered at the factory and do not need to be changed by the operator.

The properties of the leak source along with the current internal temperature of the instrument are used to compute the actual leak rate when a calibration is requested. This number is displayed as the “target” value during calibration. A spring switch located in front of the CAL Port ensures that the probe is kept in place for the duration of the calibration process. Calibration will fail if the probe is withdrawn too early in the process, or if the computed parameters deviate from the stored factory calibration by more than 20%.

1.Turn ON instrument and allow it to complete its 120 second warm-up.

2.Press the front panel CAL button to display the Calibration screen. Note that the display must be reading a leak rate of “0.00” when the CAL button is pressed; otherwise, the message “LEAK RATE MUST BE ZERO TO CALIBRATE” will appear. If necessary, press the ZERO button to zero the display before pressing CAL.

0.89TARGET Oz/yrO/YR R134A

INSERT

R134A

SETUP

MEASURED

PROBE

O/YR

INT CAL

3.Ensure that the instrument is set up for internal calibration as indicated by “INT CAL” being displayed at the bottom-right of the display. If “EXT CAL” (external calibration) is displayed, perform the procedure in Section 3.1 Selecting Internal or External Calibration and set the Cal Type to Internal.

4.Insert the probe tip into the CAL Port within 10 seconds after pressing the CAL button. Be sure that the probe tip makes contact with the spring switch. At this time the calibration process should automatically start.

The top number of the Calibration screen shows the “target” value (leak source rate), while the bottom number shows the current measured value. Keep the probe tip firmly against the back of the port for the duration of the calibration process, during which time the instrument is calibrated at both a high and low flow rate.

If the probe tip looses contact with the spring switch before the calibration process completes, the message “CAL ABORTED” is displayed. At this time press any key to retain the previous calibration curve and return to the Data Display screen.

5.After a successful calibration the new calibration data is saved and the message “CAL OK” followed by “WITHDRAW PROBE FROM LEAK SOURCE” will appear.

6.Withdrawing the probe causes the display to switch back to the Data Display screen. Note, however, if the PCHK Mode is set to AUTO (refer to Section 2.4.7 PCHK Mode), then a pressure check is performed prior to the display of the Data Display screen.

NOTE: After being calibrated the instrument is automatically placed into its Search mode.

0.89

 

 

 

 

TARGET

 

Oz/yr

 

R134AO/YR

 

R134A

 

0.00

O/YR

 

SETUP

 

 

 

MEASURED

 

 

 

 

 

 

INT CAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL

 

 

R134A

 

OK

 

 

Oz/yr

 

 

 

SETUP

 

 

 

INT CAL

 

 

 

 

 

 

 

 

 

 

 

 

WITHDRAW PROBE

Oz/yr

 

FROM LEAK

R134A

 

SOURCE

SETUP

 

 

 

INT CAL

 

 

 

 

 

 

PCHK

Oz/yr

R134A

SETUP

SM SRCH

Instruction 3015-4342

23

Image 27
Contents Infrared Refrigerant Gas Leak Detector H25-IRRegister Your Warranty by Visiting WarrantyTable of Contents Parts and Service Introduction Safety PrecautionsHazard Symbols AC Power and GroundingMisuse and Modifications to Instrument Case of Malfunction CleaningExplosive Atmosphere Working Inside the InstrumentUnpacking and Initial Checks Ordering InformationFeatures and Capabilities Leak Rate = Flow Rate x Gas Concentration Functional OverviewFront View Names and Functions of ComponentsRear View Interior Standard Probe with LED and Programmable Pushbutton Standard ProbeSmart Probe Smart Probe Keylock Function System Menu Menu SystemTurning on the Instrument Connecting the ProbeConnecting the AC Power Cord Preparing for OperationLeak Units System SetupMain Menu System Setup Menu Gas Type4 4-20 mA Loop RelaysClock Pchk ModeReset Defaults BrightnessConfiguring the Setups Function Setup NameRange Setpoint Feedback ModeSearch Auto Zero CalibrationExternal Relay Connections Typical Relay 1 Wiring4-40 mA Current Loop Connection MA Current Loop ConnectorPreparing for Operation H25-IR Calibration Data Menu CalibrationSelecting Internal or External Calibration Internal Calibration SelectionExternal Calibration Selection and Setup Internal Calibration CALExternal Calibration Procedure CAL FactorMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Restoring Factory Calibration Adjusting the CAL FactorCalibration H25-IR Operation Using the H25-IRPower Initial SetupRange Selection Setup SelectionSearching for a Leak Volume AdjustmentMeasuring the PPM Level Measuring the Leak RateLogging Measured Readings Working with Logged MeasurementsLog Menu Clearing the Log RS-232 Communication What to Do when a Fault OccursWhat is PCHK? Diagnostics MenuOperation H25-IR Maintenance Diagnostics MenuSensor Data Last Fault Factory Factory Menu Factory MenuIR Emitter PressureLeak Temp Auto GainLeak Rate Leak GasSensor Cal Leak TempcoModel Select Probe TypeFirmware Items Required Flashing MemorySelect Setup Communications Select File Load Flash Image Removing the Top Cover Part ReplacementProbe Tip Filter & O-Ring Replacement Fuse ReplacementInternal Filter Replacement Replacement Parts Parts and ServiceUnited States Service CentersCFCs, HCFCs, HFCs and Halogens SpecificationsSpecifications H25-IR Continuous Appendix a Factory DefaultsAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA