Bacharach H25-IR manual Fuse Replacement, Probe Tip Filter & O-Ring Replacement

Page 51

H25-IR

Maintenance

 

 

5.4.2 Fuse Replacement

The H25-IR is protected from electrical damage by a single 1 A fuse located next to the AC receptacle at the rear of the instrument.

Items Required:

Small flat-blade screwdriver

Fuse (P/N 604-2610 – spare fuse located in fuse holder)

Procedure:

1.Remove AC power cord from instrument.

2.Using a small flat-blade screwdriver, carefully pry the fuse holder out from the fuse compartment.

3.Visually inspect the fuse for damage or check its continuity. Replace the fuse as necessary.

5.4.3 Probe Tip Filter & O-Ring Replacement

A filter located in the probe tip is the instrument’s first line of defense against dust, dirt, and moisture from entering the instrument’s internal plumbing and IR detector. Replace this filter as it becomes dirty. A clogged filter is identified by a “Search Flow Fault,” “Measure Flow Fault,” or a “No Flow” fault occurring. Refer to Section 4.3 What to Do when a Fault Occurs.

Periodically inspect the probe tube’s O-ring for signs of cracking. If this O-ring becomes defective, air could be drawn into the instrument through the bottom of the probe tube resulting in inaccurate measurements.

Items Required:

Probe Tip Filter (P/N 3015-1571)

Probe Tube O-Ring (P/N 3015-2664)

Pin or Tweezers

Procedure:

1.Unscrew probe tip.

2.Pick out old filter with a pin or tweezers.

3.Insert new filter inside probe tip.

4.Replace probe tip.

5.Unscrew probe tube and inspect O-ring. Replace O-ring if defective.

6.Reattach probe tube.

Instruction 3015-4342

47

Image 51
Contents Infrared Refrigerant Gas Leak Detector H25-IRRegister Your Warranty by Visiting WarrantyTable of Contents Parts and Service Introduction Safety PrecautionsHazard Symbols AC Power and GroundingMisuse and Modifications to Instrument Case of Malfunction CleaningExplosive Atmosphere Working Inside the InstrumentUnpacking and Initial Checks Ordering InformationFeatures and Capabilities Leak Rate = Flow Rate x Gas Concentration Functional OverviewFront View Names and Functions of ComponentsRear View Interior Standard Probe with LED and Programmable Pushbutton Standard ProbeSmart Probe Smart Probe Keylock Function System Menu Menu SystemTurning on the Instrument Connecting the ProbeConnecting the AC Power Cord Preparing for OperationLeak Units System SetupMain Menu System Setup Menu Gas Type4 4-20 mA Loop RelaysClock Pchk ModeReset Defaults BrightnessConfiguring the Setups Function Setup NameRange Setpoint Feedback ModeSearch Auto Zero CalibrationExternal Relay Connections Typical Relay 1 Wiring4-40 mA Current Loop Connection MA Current Loop ConnectorPreparing for Operation H25-IR Calibration Data Menu CalibrationSelecting Internal or External Calibration Internal Calibration SelectionExternal Calibration Selection and Setup Internal Calibration CALExternal Calibration Procedure CAL FactorMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Restoring Factory Calibration Adjusting the CAL FactorCalibration H25-IR Operation Using the H25-IRPower Initial SetupRange Selection Setup SelectionSearching for a Leak Volume AdjustmentMeasuring the PPM Level Measuring the Leak RateLogging Measured Readings Working with Logged MeasurementsLog Menu Clearing the Log RS-232 Communication What to Do when a Fault OccursWhat is PCHK? Diagnostics MenuOperation H25-IR Maintenance Diagnostics MenuSensor Data Last Fault Factory Factory Menu Factory MenuIR Emitter PressureLeak Temp Auto GainLeak Rate Leak GasSensor Cal Leak TempcoModel Select Probe TypeFirmware Items Required Flashing MemorySelect Setup Communications Select File Load Flash Image Removing the Top Cover Part ReplacementProbe Tip Filter & O-Ring Replacement Fuse ReplacementInternal Filter Replacement Replacement Parts Parts and ServiceUnited States Service CentersCFCs, HCFCs, HFCs and Halogens SpecificationsSpecifications H25-IR Continuous Appendix a Factory DefaultsAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA