Bacharach H25-IR manual Restoring Factory Calibration, Adjusting the CAL Factor

Page 31

H25-IR

Calibration

 

 

3.4.3 Adjusting the CAL Factor

Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to highlight “CAL DATA”, and then press ENT to display the Calibration Data menu.

Main Menu:

SYSTEM

 

 

 

LOG

 

 

 

SETUPS

Calibration Data Menu:

CAL DATA

DIAGNOSTICS

 

CAL TYPE

 

 

EXT RATE

 

 

EXT GAS TYPE

 

 

CAL FACTOR

 

 

 

RESTORE FACTORY CAL

From the Calibration Data menu, highlight “CAL FACTOR” and press ENT to display its setup screen. Use the Left and Right buttons to move the blinking cursor across the screen to highlight the digit to be modified, and then use the Up and Down buttons to modify that digit. Press ENT to save the new CAL factor value and return to the Calibration Data menu.

CAL FACTOR 1.000

FOR R134A

K1=310.89 K2=703.7 K3=1460.3

K4=0.0 K5=0.0

3.5 Restoring Factory Calibration

NOTE: Factory calibration cannot be restored if the instrument was calibrated using an external leak source.

Restore an instrument to its factory calibration value as follows:

Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to highlight “CAL DATA”, and then press ENT to display the Calibration Data menu.

Main Menu:

SYSTEM

 

 

LOG

 

 

SETUPS

Calibration Data Menu:

CAL DATA

DIAGNOSTICS

 

CAL TYPE

 

 

EXT RATE

 

 

EXT GAS TYPE

 

 

CAL FACTOR

 

 

RESTORE FACTORY CAL

From the Calibration Data menu, use the Up and Down buttons to highlight “RESTORE FACTORY CAL”, and then press ENT to display the RESTORE FACTORY CAL screen.

RESTORE FACTORY CAL

HIT ENTER TO CONFIRM

Press the ENT button to confirm restoring the instrument to its factory calibration values.

Instruction 3015-4342

27

Image 31
Contents Infrared Refrigerant Gas Leak Detector H25-IRRegister Your Warranty by Visiting WarrantyTable of Contents Parts and Service Introduction Safety PrecautionsHazard Symbols AC Power and GroundingMisuse and Modifications to Instrument Case of Malfunction CleaningExplosive Atmosphere Working Inside the InstrumentUnpacking and Initial Checks Ordering InformationFeatures and Capabilities Leak Rate = Flow Rate x Gas Concentration Functional OverviewFront View Names and Functions of ComponentsRear View Interior Standard Probe with LED and Programmable Pushbutton Standard ProbeSmart Probe Smart Probe Keylock Function System Menu Menu SystemTurning on the Instrument Connecting the ProbeConnecting the AC Power Cord Preparing for OperationLeak Units System SetupMain Menu System Setup Menu Gas Type4 4-20 mA Loop RelaysClock Pchk ModeReset Defaults BrightnessSetup Name Configuring the Setups FunctionRange Setpoint Feedback ModeSearch Auto Zero CalibrationExternal Relay Connections Typical Relay 1 Wiring4-40 mA Current Loop Connection MA Current Loop ConnectorPreparing for Operation H25-IR Calibration Data Menu CalibrationSelecting Internal or External Calibration Internal Calibration SelectionExternal Calibration Selection and Setup Internal Calibration CALExternal Calibration CAL Factor ProcedureMaterial Required Improving PPM Measurement Accuracy using the CAL Factor Restoring Factory Calibration Adjusting the CAL FactorCalibration H25-IR Operation Using the H25-IRPower Initial SetupRange Selection Setup SelectionSearching for a Leak Volume AdjustmentMeasuring the PPM Level Measuring the Leak RateWorking with Logged Measurements Logging Measured ReadingsLog Menu Clearing the Log RS-232 Communication What to Do when a Fault OccursWhat is PCHK? Diagnostics MenuOperation H25-IR Diagnostics Menu MaintenanceSensor Data Last Fault Factory Factory Menu Factory MenuIR Emitter PressureLeak Temp Auto GainLeak Rate Leak GasSensor Cal Leak TempcoProbe Type Model SelectFirmware Items Required Flashing MemorySelect Setup Communications Select File Load Flash Image Removing the Top Cover Part ReplacementProbe Tip Filter & O-Ring Replacement Fuse ReplacementInternal Filter Replacement Replacement Parts Parts and ServiceUnited States Service CentersCFCs, HCFCs, HFCs and Halogens SpecificationsSpecifications H25-IR Continuous Appendix a Factory DefaultsAppendix a H25-IR H25-IR Appendix a Headquarters Hunt Valley Circle, New Kensington, PA