Rosemount 1154
Reference Manual
Process Flange Reassembly
1.Replace the metal
2.Carefully place an
3.Carefully place the flange (13 or 15) as shown in Figure
4.On differential units, repeat steps 2 and 3 for the low side (“L”) of the module. If a gage unit has two
5.Keeping the flanges parallel to each other and to the module faces, insert the four bolts (14) (and four washers on Range Code 9) and
Each spare parts kit contains the correct number of nuts, bolts, and washers for the specific transmitter range code it is designated for. Due to consolidation of parts kits, the bolt length and quantity of washers required may differ from existing transmitter assemblies and/or parts kits. Verify by part number that the appropriate spare parts kit is used for the transmitter range code being
6.Evenly seat the flanges on the sensor module housing, using a hand torque wrench as specified in steps 7 through 11. See Figure
7.Alternately tighten bolts A and B to 10
8.Alternately tighten bolts C and D to 10
9.Check the torque on bolts A and B.
10.Check the torque on bolts C and D.
11.Repeat steps
30
12.Expose all ranges of gage transmitters to two temperature cycles over the expected temperature operating range before calibrating. Expose differential and