Correction For High Line Pressure (Rosemount 1154DP and 1154HP Only)
The damping adjustment permits damping of rapid pressure variations by adjusting the
To adjust the damping, turn the damping adjustment potentiometer until the desired
The damping adjustment potentiometer has positive stops at both ends. Forcing the potentiometer beyond the stops may cause permanent damage.
NOTE
If you remove either cover during the above procedures, replace the
Span
If a differential transmitter is calibrated with the low side at ambient pressure but will be used at high line pressure, correct the span adjustment to compensate for the effect of static pressure on the unit. If zero is elevated or suppressed, also correct the zero adjustment. Correction factors, expressed in percent of differential pressure input at end points per 1,000 psi static pressure, are:
Range Codes 4, 5, and 8:
+0.75% of input/1,000 psi
Range Codes 6 and 7:
+1.25% of input/1,000 psi
The correction procedure below uses the following example:
Range Code 5, calibrated at