Emerson 1154, 00809-0100-4514 manual Specifications and Reference Data, Nuclear Specifications

Page 45

Reference Manual

00809-0100-4514, Rev BA January 2008

Rosemount 1154

Section 6 Specifications and Reference Data

NUCLEAR SPECIFICATIONS

Nuclear Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1 Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page 6-2 Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page 6-4 Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-6 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-8 Spare Parts Shelf Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-11 Important Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-11

Qualified per IEEE Std 323-1974 and IEEE Std 344-1975, as stated in Rosemount Report D8400102.

Radiation

Accuracy within ±(1.5% of upper range limit + 1.0% of span) during and after exposure to 55 megarads TID gamma radiation at the centerline at the following dose rate: 2 megarads/hr for 2 hr,

1.5 megarad/hr for 4 hr, 1 megarad/hr up to

55 megarads TID and an additional 55 megarads TID at a rate of 1 megarad/hr during post-accident operation

Range Code 0: ±(2.25% of upper range limit + 1.0% of span)

Seismic

Accuracy within ±0.5% of upper range limit after a seismic disturbance defined by a required response spectrum with a ZPA of 7 g

Range Code 0: ±0.75% of upper range limit

Steam Pressure/Temperature

Accuracy within ±(2.5% upper range limit + 0.5% of span) during and after sequential exposure to steam at the following temperatures and pressures, concurrent with chemical spray for the first 24 hr:

420 °F (215.6 °C), 50 psig for 3 minutes

350 °F (176.6 °C), 110 psig for 7 minutes

320 °F (160.0 °C), 75 psig for 8 hours

265 °F (129.4 °C), 24 psig for 56 hours

Range Code 0: ±(3.75% of upper range limit + 0.5% of span)

Chemical Spray

Composition is 0.28 molar boric acid, 0.064 molar sodium thiosulfate, and sodium hydroxide to make an initial pH of 11.0 and a subsequent pH ranging from 8.5 to 11.0. Chemical spray is sprayed at a rate of

0.25 gal/min/ft2.

www.rosemountnuclear.com

Image 45
Contents Reference Manual RosemountImportant Notice -- Errata Rosemount 1154 Alphaline Nuclear Pressure Transmitter Return of Material Revision Status Changes From June 1999 to JanuaryNew Old Changes Rosemount Table of Contents Maintenance Troubleshooting Specifications Reference DataSection TOC-3 TOC-4 Overview About the Transmitter Section IntroductionRosemount Overview General Considerations Section InstallationProcess Connections Mechanical ConsiderationsRosemount Electrical Considerations ConduitTransmitter Wiring Connections Installation ProceduresMechanical Range Code Pressure DimensionElectrical Liquid ServiceTransmitter Terminal Block Rosemount Span Adjustment Zero Adjustment Section CalibrationOverview Calibration Zero and Span Adjustment Calibration ProceduresZero and Span Adjustment Screws Linearity Adjustment Linearity and Damping Adjustment Damping AdjustmentRosemount Calculate zero adjustment correction in terms of pressure Rosemount Overview Transmitter Operation Section OperationΔ-Cell Demodulator Linearity AdjustmentFV Iref Overview Maintenance TroubleshootingTest Terminals Board Checkout Safety MessagesSensing Module Checkout Process Flange Removal Disassembly ProcedurePart Description Rosemount 1154 Parts ListReassembly Procedure Rosemount Process Flange Reassembly Post Assembly Tests Symptom Potential Source Corrective ActionOn Next Check for adequate voltage to the transmitter Nuclear Specifications Specifications and Reference DataPerformance Specifications Range Code Overpressure Effect Functional Specifications Range Code Static Pressure Zero EffectRange Code Response Time 25 to 0-150 inH2O 0-6.22 to 0-37.3 kPa Range Code Pressure RangesRange Code Overpressure Limit Physical SpecificationsRosemount Ordering Information Rosemount 1154DP 1154HP 1154GP Part Description Order No Rosemount 1154DP, 1154HP, and 1154GP Spare PartsFor use with existing N0026 transmitter only Spare Parts Shelf Life Important Notice Rosemount Current Control Current Limit Index-2 Page ¢00809-XXXX-XXXX ¤