Emerson IFT9701 instruction manual Visibility of tags, Transmitters approved for Atex Zone

Page 15

Getting Started continued

2.1.4Visibility of tags

For personal and system safety, all tags attached to the transmitter housing must remain visible. Clean or replace them as necessary.

2.2Transmitters approved for ATEX Zone 1

WARNING

Explosion hazard

The circuit board compartment is rated EEx d (flameproof), and should remain closed at all times after the flowmeter has been installed.

To avoid risk of explosion in an explosive atmosphere:

Read label that points to compartment cover before accessing circuit board compartment. Figure 2-2 shows the location of the labels on the transmitter housing.

Shut off power, and wait at least 2 minutes before removing EEx d (circuit board) compartment cover. Figure 2-2 also identifies the EEx e (field wiring), EEx i (sensor wiring), and EEx d (circuit board) compartments.

If the transmitter carries an ATEX Zone 1 approval, wiring compartments are labeled as shown in Figure 2-2.

The sensor wiring compartment is rated EEx i (intrinsically safe), and may be opened at any time. See Label 1 in Figure 2-2.

The field wiring compartment is rated EEx e (increased safety), and should remain closed when power is on. See Label 2 in Figure 2-2.

The circuit board compartment is rated EEx d (flameproof), and should remain closed at all times after the transmitter has been installed. If the transmitter is approved by ATEX as flameproof, the compartment has a lockout device, shown in Figure 2-2. The lockout device must be loosened and rotated before the compartment cover can be unscrewed.

Before You Begin

Getting Started

Mounting the Remote Transmitter

Power Supply and Output Wiring

Model IFT9701 Transmitter Instruction Manual

7

Image 15
Contents Micro Motion Page Micro Motion Page Contents Flowmeter Startup Index Page Before You Begin About this manual SafetyEuropean installations DefinitionsFlowmeter components Remotely mounted IFT9701 transmitter Page Safety, reliability, accessibility Getting StartedHazardous area Orientation and mounting Temperature, humidity, and vibrationTransmitters approved for Atex Zone Visibility of tagsCompartment tags and lockout device Jumper settingsSecurity Fault output levelsRe-installing the circuit board compartment cover Remotely Mounting the Transmitter Installation dimensions for remote mounting Mounting the transmitter to a wall or other surface Cable types Connecting the transmitter to the sensorCable selection guidelines Temperature ranges for jacket materialWiring connections to sensor Guidelines for conduitGuidelines for cable gland Sensor terminal designations Cable connections to sensorsSensor wiring compartment and sensor wiring terminals Wiring guidelines Power Supply and Output WiringPower supply wiring terminals Connect power supply wiringConnect output wiring Wire guidelines for DC power supplyMilliamp output performance Milliamp outputMilliamp output connected to Bell 202 multidrop network Typical Hart network wiringHart Communicator connections Communication tools connected to milliamp outputPulse output Wiring to pulse counter with internal pull-up resistor Wiring to Series 3000 applications platform with I/O cable Page Zeroing Customer serviceStartup procedures Flowmeter StartupInitialization Diagnostic LEDStartup mode Operating modeFlowmeter zeroing Zeroing procedureDiagnosing zero failure Configuration, calibration, and characterizationProcess measurement Page Configuration overview Configuration with a Hart CommunicatorConfiguration parameters Hart tagMeasurement units Mass flow unit Software label Volume flow unit Measurement units for mass and volume flowDamping Flow cutoffEffect of flow direction on outputs and totalizers Flow directionPulse output scaling Range values for milliamp outputAuto zero Calibration proceduresDiagnosing zeroing failure Flow calibration procedureTotal Characterization Configuration with ProLink II Software Configuration window Device panel Process variable measurement Configuration window Flow panelMass flow unit Software label Volume flow unit Effect of flow direction on outputs and totalizers Configuration window Analog Output panel Output configurationConfiguration window Frequency panel Configuration with ProLink II Software Flow Calibration dialog box Diagnosing zeroing failure Totalizer Control window Characterization Configuration window Density panel Troubleshooting Transmitter diagnostic toolsGeneral guidelines Fault outputs Optional LCDNormal conditions indicated by LED Conditions indicated by optional LCD Power supplyWiring Power supply connectionsOver range and sensor failure conditions Normal resistance and voltage ranges for flowmeter circuitsSlug flow Troubleshooting over range and sensor failure conditionsTransmitter failure Digital diagnostic messagesPage Pulse passive IFT9701 SpecificationsPerformance specifications Functional specifications Output signals Milliamp activeLocal display optional Fault indicationOutput testing Current source Power supply options 85 to 250 VACStorage Environmental limits Temperature OperatingProcess fluid vs. ambient temperature 20 to 30 VDCHumidity limits Vibration limitsDensity limits Hazardous area classifications Ambient temperature effect on milliamp outputShipping weight Environmental effectsPage Hart Communicator Menu Trees Appendix BHart Communicator Menu Trees Appendix C Installing the Optional DisplayInstalling the Optional Display Return Policy New and unused equipmentUsed equipment Page Atex IndexHart Customer service Index Page 3100572