Configuration with ProLink II Software continued
7.2Configuration parameters
Use the Configuration window to perform the following tasks:
•Assigning a HART tag to the flowmeter
•Changing process variable measurement:
-Changing measurement units for the mass flow or volume flow rate
-Changing the flow cutoff value
-Changing the internal damping value
-Changing the flow direction parameter
•Setting range values for the milliamp output
•Scaling the pulse output
7.2.1HART tag
The HART tag consists of up to eight characters that identify the flowmeter when it communicates with other devices in a HART multidrop network.
To assign a HART tag to a flowmeter:
1.From the ProLink menu, click on Configuration.
2.In the Configuration window, click on the Device tab.
3.When the Device panel appears (as depicted in Figure
Figure 7-1 Configuration window – Device panel
46 | Model IFT9701 Transmitter Instruction Manual |