Emerson IFT9701 instruction manual Cable connections to sensors, Sensor terminal designations

Page 25

Remotely Mounting the Transmitter continued

Figure 3-7 Cable connections to sensors

ELITE and F-Series

sensor terminals

Flowmeter

cable

IFT9701

terminals

Before

Green

White

Brown

Violet

Yellow

Orange

Blue

Gray

Red

Brown

Red Clip drain wire back

Green

White Clip drain wire back

Blue

Gray Clip drain wire back

Orange

Violet

Yellow Clip drain wire back

Maximum cable length 1000 ft. (330 m)

Black

(Drains from all wire sets)

Brown

Red

Green 10

White 11

12

Blue GND

Gray

Orange

Violet

Yellow

You Begin

Prepare cable in accordance with the

instructions that are shipped with the cable

D and DL

Flowmeter

sensor terminals*

cable

 

Maximum cable length 1000 ft. (300 m)

Brown

Brown

Red

Red

Clip drain wire back

Orange

Green

White

Yellow

Clip drain wire back

Green

Blue

Blue

Gray

Clip drain wire back

Violet

Orange

Gray

Violet

Yellow

White

Clip drain wire back

Black

(Drains from all wire sets)

Brown

Red

Green

White

Blue

Gray

Orange

Violet

Yellow

IFT9701

terminals

10

11

12 GND

Getting Started

Prepare cable in accordance with the

instructions that are shipped with the cable

*Model D600 and DT sensors cannot be used with IFT9701 transmitters

Table 3-3 Sensor terminal designations

Terminal number

Wire color

Function

 

 

 

1

Brown

Drive +

 

 

 

2

Red

Drive –

 

 

 

3

Orange

Temperature –

 

 

 

4

Yellow

Temperature lead length compensator

 

 

 

5

Green

Left pickoff +

 

 

 

6

Blue

Right pickoff +

 

 

 

7

Violet

Temperature +

 

 

 

8

Gray

Right pickoff –

 

 

 

9

White

Left pickoff –

 

 

 

Mounting the Remote Transmitter

Power Supply and Output Wiring

Model IFT9701 Transmitter Instruction Manual

17

Image 25
Contents Micro Motion Page Micro Motion Page Contents Flowmeter Startup Index Page Before You Begin About this manual SafetyDefinitions European installationsFlowmeter components Remotely mounted IFT9701 transmitter Page Getting Started Safety, reliability, accessibilityHazardous area Orientation and mounting Temperature, humidity, and vibrationTransmitters approved for Atex Zone Visibility of tagsCompartment tags and lockout device Jumper settingsSecurity Fault output levelsRe-installing the circuit board compartment cover Remotely Mounting the Transmitter Installation dimensions for remote mounting Mounting the transmitter to a wall or other surface Cable types Connecting the transmitter to the sensorCable selection guidelines Temperature ranges for jacket materialGuidelines for conduit Wiring connections to sensorGuidelines for cable gland Sensor terminal designations Cable connections to sensorsSensor wiring compartment and sensor wiring terminals Wiring guidelines Power Supply and Output WiringPower supply wiring terminals Connect power supply wiringConnect output wiring Wire guidelines for DC power supplyMilliamp output performance Milliamp outputMilliamp output connected to Bell 202 multidrop network Typical Hart network wiringHart Communicator connections Communication tools connected to milliamp outputPulse output Wiring to pulse counter with internal pull-up resistor Wiring to Series 3000 applications platform with I/O cable Page Startup procedures Customer serviceFlowmeter Startup ZeroingStartup mode Diagnostic LEDOperating mode InitializationFlowmeter zeroing Zeroing procedureDiagnosing zero failure Configuration, calibration, and characterizationProcess measurement Page Configuration overview Configuration with a Hart CommunicatorHart tag Configuration parametersMeasurement units Mass flow unit Software label Volume flow unit Measurement units for mass and volume flowDamping Flow cutoffEffect of flow direction on outputs and totalizers Flow directionPulse output scaling Range values for milliamp outputAuto zero Calibration proceduresDiagnosing zeroing failure Flow calibration procedureTotal Characterization Configuration with ProLink II Software Configuration window Device panel Process variable measurement Configuration window Flow panelMass flow unit Software label Volume flow unit Effect of flow direction on outputs and totalizers Configuration window Analog Output panel Output configurationConfiguration window Frequency panel Configuration with ProLink II Software Flow Calibration dialog box Diagnosing zeroing failure Totalizer Control window Characterization Configuration window Density panel Transmitter diagnostic tools TroubleshootingGeneral guidelines Optional LCD Fault outputsNormal conditions indicated by LED Conditions indicated by optional LCD Power supplyWiring Power supply connectionsOver range and sensor failure conditions Normal resistance and voltage ranges for flowmeter circuitsSlug flow Troubleshooting over range and sensor failure conditionsTransmitter failure Digital diagnostic messagesPage Performance specifications IFT9701 SpecificationsFunctional specifications Output signals Milliamp active Pulse passiveOutput testing Current source Fault indicationPower supply options 85 to 250 VAC Local display optionalProcess fluid vs. ambient temperature Environmental limits Temperature Operating20 to 30 VDC StorageVibration limits Humidity limitsDensity limits Shipping weight Ambient temperature effect on milliamp outputEnvironmental effects Hazardous area classificationsPage Hart Communicator Menu Trees Appendix BHart Communicator Menu Trees Appendix C Installing the Optional DisplayInstalling the Optional Display New and unused equipment Return PolicyUsed equipment Page Atex IndexHart Customer service Index Page 3100572