Emerson IFT9701 instruction manual Flow Calibration dialog box Diagnosing zeroing failure

Page 61

Configuration with ProLink II Software continued

Figure 7-5 Flow Calibration dialog box

Diagnosing zeroing failure

If zeroing fails:

The Calibration Failure status light turns red.

The meter’s diagnostic LED blinks ON four times per second.

The flowmeter produces fault outputs.

The message “ELEC0” blinks on and off in the meter’s optional display. The most common sources of zeroing failure are:

Flow of fluid through meter during zeroing

Partially empty flow tubes

An improperly mounted meter

Re-zero the flowmeter after correcting the problem, or cycle power to the meter to abort the entire zeroing procedure and return to the previously established zero.

7.3.2Flow calibration procedure

Flow calibration involves adjusting the flow calibration factor so it accurately represents the sensitivity of the flowmeter to mass flow. Performing the flow calibration procedure in the field is optional.

Flow calibration is performed by running a batch of fluid through the meter, weighing the fluid, then using the Totalizer Control window to compare the weighed amount with the amount of fluid measured by the flowmeter. The accuracy of the scale used for weighing the fluid will determine the accuracy of the flow calibration. Use a scale that is highly accurate.

Flowmeter Startup

Configuration with Communicator

Configuration with ProLink II

Troubleshooting

Model IFT9701 Transmitter Instruction Manual

53

Image 61
Contents Micro Motion Page Micro Motion Page Contents Flowmeter Startup Index Page Before You Begin About this manual SafetyDefinitions European installationsFlowmeter components Remotely mounted IFT9701 transmitter Page Getting Started Safety, reliability, accessibilityHazardous area Orientation and mounting Temperature, humidity, and vibrationTransmitters approved for Atex Zone Visibility of tagsCompartment tags and lockout device Jumper settingsSecurity Fault output levelsRe-installing the circuit board compartment cover Remotely Mounting the Transmitter Installation dimensions for remote mounting Mounting the transmitter to a wall or other surface Cable types Connecting the transmitter to the sensorCable selection guidelines Temperature ranges for jacket materialGuidelines for conduit Wiring connections to sensorGuidelines for cable gland Sensor terminal designations Cable connections to sensorsSensor wiring compartment and sensor wiring terminals Wiring guidelines Power Supply and Output WiringPower supply wiring terminals Connect power supply wiringConnect output wiring Wire guidelines for DC power supplyMilliamp output performance Milliamp outputMilliamp output connected to Bell 202 multidrop network Typical Hart network wiringHart Communicator connections Communication tools connected to milliamp outputPulse output Wiring to pulse counter with internal pull-up resistor Wiring to Series 3000 applications platform with I/O cable Page Startup procedures Customer serviceFlowmeter Startup ZeroingStartup mode Diagnostic LEDOperating mode InitializationFlowmeter zeroing Zeroing procedureDiagnosing zero failure Configuration, calibration, and characterizationProcess measurement Page Configuration overview Configuration with a Hart CommunicatorHart tag Configuration parametersMeasurement units Mass flow unit Software label Volume flow unit Measurement units for mass and volume flowDamping Flow cutoffEffect of flow direction on outputs and totalizers Flow directionPulse output scaling Range values for milliamp outputAuto zero Calibration proceduresDiagnosing zeroing failure Flow calibration procedureTotal Characterization Configuration with ProLink II Software Configuration window Device panel Process variable measurement Configuration window Flow panelMass flow unit Software label Volume flow unit Effect of flow direction on outputs and totalizers Configuration window Analog Output panel Output configurationConfiguration window Frequency panel Configuration with ProLink II Software Flow Calibration dialog box Diagnosing zeroing failure Totalizer Control window Characterization Configuration window Density panel Transmitter diagnostic tools TroubleshootingGeneral guidelines Optional LCD Fault outputsNormal conditions indicated by LED Conditions indicated by optional LCD Power supplyWiring Power supply connectionsOver range and sensor failure conditions Normal resistance and voltage ranges for flowmeter circuitsSlug flow Troubleshooting over range and sensor failure conditionsTransmitter failure Digital diagnostic messagesPage Performance specifications IFT9701 SpecificationsFunctional specifications Output signals Milliamp active Pulse passiveOutput testing Current source Fault indicationPower supply options 85 to 250 VAC Local display optionalProcess fluid vs. ambient temperature Environmental limits Temperature Operating20 to 30 VDC StorageVibration limits Humidity limitsDensity limits Shipping weight Ambient temperature effect on milliamp outputEnvironmental effects Hazardous area classificationsPage Hart Communicator Menu Trees Appendix BHart Communicator Menu Trees Appendix C Installing the Optional DisplayInstalling the Optional Display New and unused equipment Return PolicyUsed equipment Page Atex IndexHart Customer service Index Page 3100572