Chapter 8
Troubleshooting
8.1Customer service
For assistance, phone the Micro Motion Customer Service Department. Phone numbers are listed on the title page of this manual.
8.2General guidelines
Troubleshooting a Micro Motion flowmeter is performed in two parts:
•Tests of wiring circuit integrity.
•Observation of the transmitter's diagnostic tools, which include the diagnostic LED, fault output levels, and optional LCD.
CAUTION
During troubleshooting, the transmitter could produce inaccurate flow signals.
Set control devices for manual operation while troubleshooting the flowmetering system.
Follow these general guidelines while troubleshooting a Micro Motion flowmeter:
•Before beginning the diagnostic process, become familiar with this instruction manual and with the instruction manual for the sensor.
•While troubleshooting a problem, leave the sensor in place, if possible. Problems may result from the specific environment in which the sensor operates.
•Check all signals under both flow and
8.3Transmitter diagnostic tools
In some situations, troubleshooting requires use of the transmitter's diagnostic tools, which include the diagnostic LED, fault output levels, and optional LCD.
8.3.1Diagnostic LED
The diagnostic LED indicates the operating condition of the transmitter. Table
Flowmeter Startup
Configuration with Communicator
Configuration with ProLink II
Troubleshooting
Model IFT9701 Transmitter Instruction Manual | 57 |