Emerson IFT9701 instruction manual Total

Page 51

Configuration with a HART Communicator continued

To calculate the flow calibration factor:

1.Select a mass flow unit, and send the unit to the flowmeter memory. See Table 6-1.

2.To enter the flow calibration factor into flowmeter memory, refer to the menu below and follow these steps:

1Process variables

2Diag/service

3Basic setup

4

DETAILED SETUP

 

1

CHARIZE SENSOR

 

1

FLOW CAL

 

 

5

Review

2

Config fld dev var

2

Density cal factr

3Config outputs

4Device information

a.At the online menu, choose 4 (detailed setup).

b.At the detailed setup menu, choose 1 (characterize sensor).

c.At the characterize sensor menu, choose 1 (flow calibration factor).

d.Enter the 8-digit flow calibration factor listed on the flowmeter serial number tag. The flow calibration factor can also be found on the certificate that is shipped with the meter.

e.When the display returns to the characterize sensor menu, press F2 (send) to send the factor to the flowmeter memory.

3.Zero the flowmeter (see Section 6.3.1).

4.Refer to the menu below, and follow these steps:

 

 

 

1

View fld dev vars

 

 

 

 

 

 

2

View output vars

 

 

 

 

 

 

3

View status

 

1

Totl

 

 

 

 

1

PROCESS VARIABLES

 

4

TOTALIZER CNTRL

 

 

 

2

Reset totalizer

2

Diag/Service

 

 

 

 

 

 

 

 

 

 

3

Basic setup

 

 

 

 

 

 

4

Detailed setup

 

 

 

 

 

 

5

Review

 

 

 

 

 

 

 

 

 

 

 

 

 

 

a.Reset the internal totalizer to zero.

b.Run three batches of fluid, resetting the scale and totalizer between batches. For each batch, record the weights indicated by the scale and the totalizer.

WeightscaleWeighttotalizer

First batch

Second batch

Third batch

Total

5.Divide Total Weightscale by Total Weighttotalizer. This is the mass-flow meter factor. Record the meter factor.

Mass-flow meter factor

Flowmeter Startup

Configuration with Communicator

Configuration with ProLink II

Troubleshooting

Model IFT9701 Transmitter Instruction Manual

43

Image 51
Contents Micro Motion Page Micro Motion Page Contents Flowmeter Startup Index Page Before You Begin About this manual SafetyEuropean installations DefinitionsFlowmeter components Remotely mounted IFT9701 transmitter Page Safety, reliability, accessibility Getting StartedHazardous area Orientation and mounting Temperature, humidity, and vibrationTransmitters approved for Atex Zone Visibility of tagsCompartment tags and lockout device Jumper settingsSecurity Fault output levelsRe-installing the circuit board compartment cover Remotely Mounting the Transmitter Installation dimensions for remote mounting Mounting the transmitter to a wall or other surface Cable types Connecting the transmitter to the sensorCable selection guidelines Temperature ranges for jacket materialWiring connections to sensor Guidelines for conduitGuidelines for cable gland Sensor terminal designations Cable connections to sensorsSensor wiring compartment and sensor wiring terminals Wiring guidelines Power Supply and Output WiringPower supply wiring terminals Connect power supply wiringConnect output wiring Wire guidelines for DC power supplyMilliamp output performance Milliamp outputMilliamp output connected to Bell 202 multidrop network Typical Hart network wiringHart Communicator connections Communication tools connected to milliamp outputPulse output Wiring to pulse counter with internal pull-up resistor Wiring to Series 3000 applications platform with I/O cable Page Zeroing Customer serviceStartup procedures Flowmeter StartupInitialization Diagnostic LEDStartup mode Operating modeFlowmeter zeroing Zeroing procedureDiagnosing zero failure Configuration, calibration, and characterizationProcess measurement Page Configuration overview Configuration with a Hart CommunicatorConfiguration parameters Hart tagMeasurement units Mass flow unit Software label Volume flow unit Measurement units for mass and volume flowDamping Flow cutoffEffect of flow direction on outputs and totalizers Flow directionPulse output scaling Range values for milliamp outputAuto zero Calibration proceduresDiagnosing zeroing failure Flow calibration procedureTotal Characterization Configuration with ProLink II Software Configuration window Device panel Process variable measurement Configuration window Flow panelMass flow unit Software label Volume flow unit Effect of flow direction on outputs and totalizers Configuration window Analog Output panel Output configurationConfiguration window Frequency panel Configuration with ProLink II Software Flow Calibration dialog box Diagnosing zeroing failure Totalizer Control window Characterization Configuration window Density panel Troubleshooting Transmitter diagnostic toolsGeneral guidelines Fault outputs Optional LCDNormal conditions indicated by LED Conditions indicated by optional LCD Power supplyWiring Power supply connectionsOver range and sensor failure conditions Normal resistance and voltage ranges for flowmeter circuitsSlug flow Troubleshooting over range and sensor failure conditionsTransmitter failure Digital diagnostic messagesPage Pulse passive IFT9701 SpecificationsPerformance specifications Functional specifications Output signals Milliamp activeLocal display optional Fault indicationOutput testing Current source Power supply options 85 to 250 VACStorage Environmental limits Temperature OperatingProcess fluid vs. ambient temperature 20 to 30 VDCHumidity limits Vibration limitsDensity limits Hazardous area classifications Ambient temperature effect on milliamp outputShipping weight Environmental effectsPage Hart Communicator Menu Trees Appendix BHart Communicator Menu Trees Appendix C Installing the Optional DisplayInstalling the Optional Display Return Policy New and unused equipmentUsed equipment Page Atex IndexHart Customer service Index Page 3100572