Emerson IFT9701 Local display optional, Low-flow cutoff, Slug-flow inhibit, Fault indication

Page 74

IFT9701 Specifications continued

A.2.2 Local display (optional)

5-digit, alphanumeric, liquid crystal display (LCD) installed on the field-wiring compartment cover. The LCD shows flow rate, and indicates slug flow, flowmeter zeroing in progress, and electronic faults.

Note: Not available for ATEX Zone 1 applications.

A.2.3 Low-flow cutoff

Flow rate below cutoff causes outputs to default to level that indicates zero flow and totalizer stops counting

A.2.4 Slug-flow inhibit

When transmitter senses density outside user-selected limits, outputs default to levels indicating zero flow and totalizer stops counting

A.2.5 Damping

User-selected time constant of 0.1, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, or 12.8 seconds

A.2.6 Fault indication

Fault condition can be indicated by user-selected downscale (2 mA, 0 Hz) or upscale (22 mA, 7200 Hz) output levels, Bell 202 communications, and optional display

A.2.7 Output testing

Current source

Can output a specified current between 2 and 22 mA

Frequency source

Can output a specified frequency between 0.01 and 1200 Hz

A.2.8 Power supply options

85 to 250 VAC

45 to 65 Hz, 9 watts maximum power dissipation

20 VA maximum apparent line power

Fused at 250 V/0.4 A IEC delayed, nonreplaceable

Meets low-voltage directive 73/23/EEC

66

Model IFT9701 Transmitter Instruction Manual

Image 74
Contents Micro Motion Page Micro Motion Page Contents Flowmeter Startup Index Page About this manual Safety Before You BeginFlowmeter components European installationsDefinitions Remotely mounted IFT9701 transmitter Page Hazardous area Safety, reliability, accessibilityGetting Started Temperature, humidity, and vibration Orientation and mountingVisibility of tags Transmitters approved for Atex ZoneJumper settings Compartment tags and lockout deviceFault output levels SecurityRe-installing the circuit board compartment cover Remotely Mounting the Transmitter Installation dimensions for remote mounting Mounting the transmitter to a wall or other surface Connecting the transmitter to the sensor Cable typesTemperature ranges for jacket material Cable selection guidelinesGuidelines for cable gland Wiring connections to sensorGuidelines for conduit Cable connections to sensors Sensor terminal designationsSensor wiring compartment and sensor wiring terminals Power Supply and Output Wiring Wiring guidelinesConnect power supply wiring Power supply wiring terminalsWire guidelines for DC power supply Connect output wiringMilliamp output Milliamp output performanceTypical Hart network wiring Milliamp output connected to Bell 202 multidrop networkCommunication tools connected to milliamp output Hart Communicator connectionsPulse output Wiring to pulse counter with internal pull-up resistor Wiring to Series 3000 applications platform with I/O cable Page Flowmeter Startup Customer serviceStartup procedures ZeroingOperating mode Diagnostic LEDStartup mode InitializationZeroing procedure Flowmeter zeroingConfiguration, calibration, and characterization Diagnosing zero failureProcess measurement Page Configuration with a Hart Communicator Configuration overviewMeasurement units Configuration parametersHart tag Measurement units for mass and volume flow Mass flow unit Software label Volume flow unitFlow cutoff DampingFlow direction Effect of flow direction on outputs and totalizersRange values for milliamp output Pulse output scalingCalibration procedures Auto zeroFlow calibration procedure Diagnosing zeroing failureTotal Characterization Configuration with ProLink II Software Configuration window Device panel Configuration window Flow panel Process variable measurementMass flow unit Software label Volume flow unit Effect of flow direction on outputs and totalizers Output configuration Configuration window Analog Output panelConfiguration window Frequency panel Configuration with ProLink II Software Flow Calibration dialog box Diagnosing zeroing failure Totalizer Control window Characterization Configuration window Density panel General guidelines TroubleshootingTransmitter diagnostic tools Normal conditions indicated by LED Fault outputsOptional LCD Power supply Conditions indicated by optional LCDPower supply connections WiringNormal resistance and voltage ranges for flowmeter circuits Over range and sensor failure conditionsTroubleshooting over range and sensor failure conditions Slug flowDigital diagnostic messages Transmitter failurePage Functional specifications Output signals Milliamp active IFT9701 SpecificationsPerformance specifications Pulse passivePower supply options 85 to 250 VAC Fault indicationOutput testing Current source Local display optional20 to 30 VDC Environmental limits Temperature OperatingProcess fluid vs. ambient temperature StorageDensity limits Humidity limitsVibration limits Environmental effects Ambient temperature effect on milliamp outputShipping weight Hazardous area classificationsPage Appendix B Hart Communicator Menu TreesHart Communicator Menu Trees Installing the Optional Display Appendix CInstalling the Optional Display Used equipment Return PolicyNew and unused equipment Page Index AtexHart Customer service Index Page 3100572