5-5-2 Using The Statistical Calculation Mode

1 Press the RE-ZERO key to set the display to zero.

2Place a container on the weighing pan.

Press the PRINT key to cancel the weight (tare). The

balance displays 0.00 g. (Storing the tare value) The tare value data is output when the peripheral output equipment is connected.

3 Weigh formula 1 and press the PRINT key. The balance displays 0.00 g. (Storing the weight value of formula 1)

The weight value data is output when the peripheral output equipment is connected.

4 Weigh formula 2 and press the PRINT key. The balance displays 0.00 g. (Storing the weight value of formula 2)

The weight value data is output when the peripheral output equipment is connected.

5When there are some more formulae to be added, repeat step 4.

6After mixing is complete, press the MODE key to display the statistical results.

7Press the PRINT key to output the number of data saved including the tare value and the total weight.

Output example