9-5 Description of the Item “Data Output Mode”

The parameter setting of the “Data output mode ( prt )” applies to the performance when the data is transmitted using the RS-232C serial interface.

Key mode

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the weight data and the display blinks one time.

Required setting dout prt 0 Key mode

Auto print modes A and B

When the displayed value is stable and the conditions of “Auto print polarity”, “Auto print difference” and reference value are met, the balance outputs the weight data.

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the data and the display blinks one time.

Mode A: Required setting

dout

prt 1

Auto print mode A (reference = zero)

 

dout

ap-p

Auto print polarity

 

dout

ap-b

Auto print difference

Example

“For outputting the weight value each time a sample is added, with

 

ar-d set to “1” (to adjust zero after the data is output).”

Mode B: Required setting

dout

prt 2

Auto print mode B (reference = last stable value)

 

dout

ap-p

Auto print polarity

 

dout

ap-b

Auto print difference

Example

“For outputting the weight value while a sample is added.”

Stream mode

The balance outputs the weight data continuously regardless of the display condition. The display does not blink in this mode.

Required setting

dout

prt 3

Stream mode

 

ba5fnc

5pd

Display refresh rate

 

5if

bp5

Baud rate

Example

“For monitoring data on a computer”

Caution

The balance may not transmit the data completely at the specified refresh rate and baud rate. Set the baud rate higher.

Key mode B

When the PRINT key is pressed, the balance outputs the weight data, regardless of the display condition. The display does not blink in this mode.

Key mode C

When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on and outputs the data. The balance blinks one time in this mode.

36

Page 38
Image 38