12. PROGRAMMABLE-UNIT

This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary coefficient set in the function table and displays the result.

The coefficient must be within the range between the minimum and maximum shown below. If the coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.

Model

Minimum coefficient

Maximum coefficient

FX-120i /200i /300i

0.000001

1000

FX-1200i /2000i /3000i

100

 

Operation

1 Press and hold the SAMPLE key until

ba5fnc

of the function table is displayed.

2

Press the

SAMPLE

key several times to display

mltj

.

 

 

 

3

Press the

PRINT

key. The balance enters the mode to confirm or set the coefficient.

Confirming the coefficient

4The current coefficient is displayed with the first digit blinking.

￿ When it is not to be changed, press the CAL key and proceed to step 6.

￿ When it is to be changed, press the RE-ZERO key and proceed to step 5.

Setting the coefficient

5 Set the coefficient using the following keys.

SAMPLE

key

To select a digit to change the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

value. The selected digit blinks.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RE-ZERO

key

To change the value.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MODE

key

To change the decimal point

 

 

 

 

 

position.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Each time the switch is pressed,

 

 

 

 

 

the decimal

 

point position

 

 

 

 

 

changes as follows:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PRINT

key

To store the new setting,

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

display

end

and go to step 6.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CAL

key

To cancel the new setting and

 

 

 

 

 

go to step 6.

 

 

 

 

 

 

 

 

 

 

 

 

 

Quitting the operation

6The balance displays Unit . Press the CAL key to exit the programmable-unit function and return to the weighing mode.

Using the function

Press the MODE key to select the programmable-unit (no display on the unit section). Perform weighing as described in “5-1 Basic Operation (Gram Mode)”. After weighing, the balance displays the result (weight data in grams x coefficient).