A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i instruction manual Glp1

Models: FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i

1 72
Download 72 pages 26.17 Kb
Page 29
Image 29

5Confirm that there is nothing on the pan and press the PRINT key. The balance measures the zero point and displays the measured value. Do not allow vibration or drafts to affect the balance.

The balance displays the target weight value.

6Place the displayed target weight on the pan and press the PRINT key. The balance measures the target weight and displays the measured value. Do not allow vibration or drafts to affect the balance.

7The balance displays pan.

end . Remove the weight from the

8When the GLP output is set, the balance displays and outputs “Calibration Test Report”.

glp1

9The balance will automatically return to the weighing mode.

27

Page 29
Image 29
A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i instruction manual Glp1