7-2 Calibration Using an External Weight

This function calibrates the balance using an external weight.

Operation

1Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the pan.

2Press and hold the CAL key until displayed, and then release the key.

Calout

is

3 The balance displays

Cal 0 .

If you want to change the calibration weight (a list of usable weights is shown on page 23), press the SAMPLE key and proceed to step 4.

If you use the calibration weight value stored in the balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE

key

To switch the display condition

 

 

 

 

 

to:

“All

of

the

segments

 

 

 

 

 

blinking”

(calibration

weight

 

 

 

 

 

selection mode) or “The last

 

 

 

 

 

two

digits

blinking”

(value

 

 

 

 

 

adjustment mode).

 

 

 

 

 

 

 

 

 

 

RE-ZERO

key (To increase the value)

 

 

 

 

 

 

 

 

 

 

MODE

key (To decrease the value)

 

 

 

 

 

 

 

 

 

To

select

the

 

 

calibration

 

 

 

 

 

weight or adjust the value.

 

 

 

 

 

 

 

 

 

PRINT

key

To

store

the

new

weight

 

 

 

 

 

value. Even

if

the AC

 

 

 

 

 

adapter is removed, the data

 

 

 

 

 

is maintained in non-volatile

 

 

 

 

 

memory.

 

 

 

 

 

 

 

 

 

 

 

CAL

key

To cancel the operation and

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

return to

Cal

0

.

 

 

24

Page 26