7-3 Calibration Test Using an External Weight

This function tests the balance weighing accuracy using an external weight and outputs the result. This is available only when the “GLP output (info)” parameter is set to “1” or “2”. (Calibration test does not perform calibration.)

Operation

1 Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until displayed, and then release the key.

CCout

is

3 The balance displays CC 0 .

If you want to change the target weight (a list of usable weights is shown on page 23), press the SAMPLE key and proceed to step 4.

If you use the target weight value stored in the balance, proceed to step 5.

4 Specify the target weight value as follows:

SAMPLE key To switch the display condition to: “All of the segments blinking” (target weight selection mode) or “The last two digits blinking” (value adjustment mode).

RE-ZERO key (To increase the value)

MODE key (To decrease the value)

To select the target weight or adjust the value.

PRINT

key

To store

the

new weight

 

 

 

value. Even

if the AC

 

 

 

adapter is removed, the data

 

 

 

is maintained in non-volatile

 

 

 

memory.

 

 

 

 

 

 

 

 

CAL

key

To cancel the operation and

 

 

 

 

 

 

 

 

 

 

return to

CC

0

.

26

Page 28
Image 28
A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i instruction manual Calibration Test Using an External Weight